MANUAL THERMOCOAGULATION - EVRF
Date
26/06/2018
Page 25 of 34
Revision N° : 18
The EVRF will give a sound beep every two seconds, so you know that every beep, you
have generated 2 times the watt that was set on the EVRF. This is the amount of joule that
the EVRF generated.
When pressing the “OFF” button the
device will check if the output cable is
connected. If the cable is not connected a
beep is given and the device remains in
“OFF” mode.
If the output cable is connected, then a
pop-up screen is shown with the status of
the pad. When you accept the pop-up then
the device will be place in “ON” mode.
If the pop-up is declined, then the device
remains in “OFF” mode.
You can also leave the program and return
to the program selection by pressing
.
5.2.4 Hemorrhoids
-
The Power setting, standard on
25Watt, can be decreased
or
increased
between 20W and
25W.
Continuous mode. When the pedal is
pressed the output is active until the
pedal is released.
Joule:
-This is the sum of the amount of
energy generated over a period of
time.
The EVRF will give a sound beep
every two seconds, so you know that
every beep, you have generated 2 times the watt that was set on the EVRF. This is the
amount of joule that the EVRF generated.
When pressing the “OFF” button the device will check if the output cable is connected. If the
cable is not connected a beep is given and the device remains in “OFF” mode.
The device will also check if the pad is connected. If the pad is connected, then a warning
will be shown advising to disconnect the pad. If the pad is not connected, then no message is
shown and the device is placed in “ON” mode.
You can also leave the program and return to the program selection by pressing . End of
the care.
To stop the care definitively, one touches the
button. We return to the basic screen.