![F Care Systems EVRF Скачать руководство пользователя страница 21](http://html.mh-extra.com/html/f-care-systems/evrf/evrf_manual_534565021.webp)
MANUAL THERMOCOAGULATION - EVRF
Date
26/06/2018
Page 21 of 34
Revision N° : 18
5
USE OF THE EVRF SYSTEM
The use of the EVRF is very easy.
Connect the foot pedal to the apparatus (blue plug).
If needed connect the non-sterile output cable to the device (only required for needle
holder CR30KAB treatment).
Switch on the apparatus, the welcome screen appears.
The welcome screen is shown while the
EVRF is testing the different programs.
In case a problem is found with one of the
possible programs a warning screen is
shown (see chapter 5.3).
If all programs are functioning correctly, then the program selection screen is shown.
In this menu you have the option to choose
between 4 modes:
-Telangiectasia with needle holder < 1mm.
-Varicose veins with CR30KAB catheter < 1-
5 mm.
-GSV/SSV with CR45i catheter.
-Haemorrhoids with HPR45i probe.
Changing the selection can be done by
touching the button.
By touching you go to the settings screen.