background image

< Programming the Setpoints and the Deadband > 

1.  Press 

  to  choose pH or  ORP mode. The  controller  will  store  the  settings for pH 

and ORP separately. 

2.  Press and hold 

  button until a beep sound, and then press 

  to enter setting 

mode. Use 

  and 

  to adjust settings, and use 

  to store each setting. 

3.  The  display  will  then  automatically  cycle  through  5  settings:  rl(relay)  on  or  off,  HIr(HI 

relay setpoint), Lor(LO relay setpoint), dbH(HI Deadband), and dbL(LO Deadband). 

4.  When setting HI or LO setpoints, press and hold 

  to switch the value to Positive or 

to Negative. 

 

 

 

Choose pH or ORP 

Enter setting mode 

HIr(HI relay setpoint)

 

Lor(LO relay setpoint)

 

dbH(HI Deadband)

 

dbL(LO Deadband)

 

End 

ON

 

OFF

 

Relay ON or OFF

 

pH: -2.00~16.00 pH 

ORP: ±1999 mV 

pH: 0.00~2.00 pH 

ORP: 0~200 mV 

Содержание 4801P

Страница 1: ... mV 500 1999 mV 0 110 C Accuracy 0 01 1 digit 2 1 digit 0 2 1 digit Resolution 0 01 pH 0 1 1 mV 0 1 C Compensation ATC 0 100 C N A Deadband pH 0 00 to 2 00 pH ORP 0 to 200 mV Temperature Compensation ATC 0 100o C via 30K Thermistor or MTC Limit Relays 2 SPDT contacts rated for 5A at 115V Relay status indicators Front panel LED lamp Power supply 9V DC Panel Cutout 1 16th DIN 1 81 x 1 81 46 x 46mm M...

Страница 2: ...l the bezel is flush with the panel 3 Slide the mounting bracket over the rear of the controller and press snugly against the rear of the panel Rear panel connection Connect the alarm control devices to the HI or LO relay terminals The SPDT relays are dry contact type and the terminal designations are C Common NO Normally Open NC Normally Closed ...

Страница 3: ...ensation ATC or Manual temperature compensation MTC 7 Temperature reading and unit 8 unit of pH or ORP reading 1 Turn on off the controller 2 Enter calibration mode 1 Switch pH and ORP mode 2 Press and hold to enter setting mode 3 In setting mode press to store each setting 1 In measurement mode adjust temperature value of MTC 2 In setting mode adjust each parameter 1 2 3 4 8 7 6 5 ...

Страница 4: ...rode must be replaced A slope of 100 is ideal Note 1 Calibration error indicator icon will appear and Err instead of SA if calibration fails 2 When doing a 2 or 3 points calibration Calibrate with buffer pH 7 first and then follow with buffer pH 4 or pH 10 pH Calibration type and temperature unit setting This controller has two pH calibration type USA and NIST and the calibration points are follow...

Страница 5: ...ection HI and LO Deadband The Deadband feature is used to minimize relay chatter rapid on off relay switching near setpoint This is accomplished by creating a region where no relay action can occur The High Deadband parameter creates a band at the high setpoint and the Low Deadband parameter creates a band at the low setpoint HI relay ON reading HI relay setpoint HI deadband HI relay OFF reading H...

Страница 6: ...each setting 3 The display will then automatically cycle through 5 settings rl relay on or off HIr HI relay setpoint Lor LO relay setpoint dbH HI Deadband and dbL LO Deadband 4 When setting HI or LO setpoints press and hold to switch the value to Positive or to Negative Choose pH or ORP Enter setting mode HIr HI relay setpoint Lor LO relay setpoint dbH HI Deadband dbL LO Deadband End ON OFF Relay ...

Страница 7: ...ht is always on OFF back light is always off Auto back light will be on for 30 seconds while any button is pressed and then off Reset the controller to default setting 1 Press and hold until a beep sound release and then press and hold again until a beep sound 2 Press to switch OFF to On and then press to reset the controller to default setting ...