4. Press the chair tightly till it fully closed as in Figure M3.
XII.Troubleshooting
The control panel will display a number of possible trouble conditions by means
of the LED light.When the LED light is green, the chair is trouble-free. When a
problem occurs, the LED light will flash.
When a “Trouble” LED light flashes, turn off the power of the chair and turn it on
again. If the flashing “Trouble” LED light remains unchanged, please refer to the
table below for possible causes and solutions.
If you cannot determine the cause of the trouble condition, please contact your
provider or distributor for assistance.
26
No. Trouble
Possible cause
Solution
1
Power
Faulty connection of the
control panel and the motor
connector
Re-plug the 4-pin plug into
the battery connector at the
side of the chair.
Low power supply, the
power cut off automatically.
If the power is off
automatically, the batteries
must be recharged within
two hours and charged for
10 – 12 hours.
Fault of the 3-pin plug
connecting to the battery.
Pull out the battery and
check the contact for any
fault and reinsert the
battery.