Brake Adjustment
(See Diagram 4)
•
There are two brake adjusters on each side of the rollator, one in
front of the brake handle and the second is positioned on the rear
legs behind the wheels.
•
To adjust the handle brake, loosen the lock nut (you may need a
tool to carry out this operation), turn the adjuster anti-clockwise to
tighten the brake then re-tighten the lock nut (diagram 4).
•
To adjust the rear legs brake, loosen the lock nut (you may need a
tool to carry out this operation), turn the adjuster counter-clockwise
to tighten the brake then retighten the lock nut. For correct
adjustment the brake mechanism, should not touch the wheel but
the gap should be no greater than3 to 4 mm.
•
Caution: After making any adjustments to the brakes, you must test
them before the rollator is used.
•
If adjustment does not tighten the brakes, do not use the rollator and
contact your dealer.
Imported / Distributed by:
Factory Direct Medical
34 Futurity Gate, Unit # 15
Concord Ontario L4K 1S6
Local: 416-739-8393
Toll Free: 855-235-2400