EZ Way, Inc.
EZ Way, Inc. • (800) 627-8940 • Fax (712) 542-1899 • www.ezlifts.com
12
Advanced Smart Stand
®
Features
ADVANCED DIAGNOSTICS
Real Time Current – For Actuator Diagnostics
1. When the display shows “EZ Stand Awake”, press the hidden key. Now whenever the stand is in
motion, the real time current will be shown.
2. The Real Time Current function will disable after power cycling the unit.
Real Time Voltage – For Battery Diagnostics
1. In the case of a possible battery or connection issue the battery voltage can be monitored real-time.
Go to the Customer Settings menu.
2. Press the Enter key several times until you see “Volts = …”.
CUSTOMER SETTINGS MENU (FULL DESCRIPTIONS)
Within the Customer Settings Menu there are several functions that provide customer feedback.
The functions include; an Inspection Reminder, stand usage, and diagnostic information. Below is a
description of each function within the Customer Settings Menu.
Inspection Reminder – For Scheduled Maintenance
The Inspection Reminder function tracks the last inspection occurrence. When the duration of time
since the last inspection exceeds 30 days a reminder message will appear when the unit is powered
up. The purpose of the inspection reminder is to notify the customer that the lift or stand needs to be
inspected with the ‘EZ Way Safety & Maintenance Checklist’ form (see attached). When the checklist is
complete the reminder is to be reset in the Customer Settings Menu as described below.
1. To enable/disable the Inspection Reminder enter the Customer Settings Menu.
2. Press enter. Scroll to Inspection Reminder. Press enter. Scroll to choose Yes or No. Press Enter.
3. Press Exit twice.
Inspection Complete
1. If ‘INSPECTION OVERDUE’ appears during power up complete the ‘EZ Way Safety &
Maintenance Checklist’.
2. Enter the Customer Settings Menu.
3. Press enter. Scroll to Inspection Complete. Press enter. Scroll to choose Yes. Press Enter.
4. Press Exit twice.
Overweight Counts – For Diagnostic Tracking with Scale
The Overweight Counts function tracks ‘Weight Overload’ occurrences. When a ‘WEIGHT
OVERLOAD’ occurs the ‘Overweight Counts’ register increments one. This read-only information
provides customer feedback for; accidental lifting of a trapped harness, or accidental lifting of overweight
patient. This information is available for units with the scale option only.