5-1-2 Attaching cyclone
1) Fit the packing (SC) into the groove of the
connector of the evaporating tube.
2) Connect connectors of the cyclone and evaporating
tube, screw the connectors and fasten the
cyclone on the evaporating tube.
3) Insert the connection collar in the cyclone.
*When insertion becomes loose, replace two O rings
inside the connection collar.
Cyclone
Connector
Evaporator
Cyclone
5-1-3 Attaching outlet temp. sensor
1) Connect the connecting portion of outlet
temperature sensor to the connector on the main
unit.
2) Slightly loosen the cap on the sensor mounting port
of the evaporating tube and insert the outlet
temperature sensor.
3) Make adjustments so that the tip of the outlet
temperature sensor is positioned in the branched
tube center of the evaporating tube and then
fasten the cap to fix the sensor.
*Be careful not to touch the cable of the outlet temp.
sensor to the evaporating tube.
②
Fasten
Connector
Sensor for outlet
temperature
Branch tube of evaporating tube
Cap
③
Fasten
-12-
Insert
Connection collar
①
Connect
Packing (SC)
②
Insert
①
Connect
Code No.
Part name
Remarks
275510
O ring for connection
collar