3-Phase Rotation 

The 3-phase rotation function indicates

 whether 

the 3 phase

 AC lines or mains 

are live and also 

identifies the phase orientation of the three wires before attaching them to a motor or other equipment. 
The phase sequence will determine the direction (Clockwise or Counter-clockwise) a motor will

 rotate 

when connected.  
 
 
To determine clockwise rotation (P1-P2-P3)  
 

1. 

Touch probe L1 to

 any one of the three AC mains. L

abel this phase P1.  

2. 

Touch probe L2 to the other two phases. One phase will indicate “L” and one will indicate “R” 

3. 

Label the phase that indicates “R” as P2. 

4. 

Touch probe L1 to phase P2 and touch probe L2 to the remaining unlabeled

 AC main 

wire. 

The “R” icon will illuminate. Label this phase P3. 

5. 

The clockwise sequence P1-P2-P3 (also known as 1-2-3 or R-S-T) has been identified. 

6. 

To determine a counterclockwise rotation exchange the “L” and “R” in the procedure. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 

Continuity Test

 

The tester can measure resistance and alert the user if the resistance value is less than 200k

Ω

1. 

Before taking a continuity test, make sure that power to the device or circuit under test is off 
and that all capacitors are discharged. 

2. 

Touch the test tips to the device under test. 

3. 

If the resistance is less than 200k

Ω

, the tester will sound an audible alert; the LCD will display 

“000”, and the  continuity LED will light. 

 
Flashlight

 

Press and hold the 

 button to turn the flashlight on. Release the button to turn the flashlight off. After 

the button is released, the light will turn off after 12 seconds (approx). 

P1 P2 P3

L1

L2

P1 P2 P3

L1

L2

L1

Содержание VT30-E

Страница 1: ...User s Guide LCD Multifunction Voltage Tester Model VT30 E E ...

Страница 2: ...use 3 Inspect the condition of the test leads and the meter itself for any damage before use 4 Do not exceed the rated input limits 5 Use great care when taking measurements greater than 25VAC rms or 35VDC These voltages are considered a shock hazard 6 Discharge capacitors and remove power from the DUT before Continuity tests 7 Remove the batteries from the meter if the meter is to be stored for l...

Страница 3: ...4 LCD display 5 LED AC voltage indicator between 100 to 480V 6 LED low impedance test 7 LED continuity 8 Low impedance test button L2 9 Flashlight button 10 Battery cover 11 Low impedance test button L1 Probe Tip Category 1 2 3 4 5 6 7 8 9 10 1 1 CAT IV 600V CAT III 600V CAT II 600V ...

Страница 4: ...ion To check for the presence of voltage between 100V and 480V using only the positive test lead L2 touch the lead to the device or circuit under test If voltage is present the LED will light Note In this mode the actual voltage is not displayed only the presence of voltage is detected Low Impedance Voltage Measurement Due to the lowered internal impedance capacitive voltage is suppressed so that ...

Страница 5: ...AC main wire The R icon will illuminate Label this phase P3 5 The clockwise sequence P1 P2 P3 also known as 1 2 3 or R S T has been identified 6 To determine a counterclockwise rotation exchange the L and R in the procedure Continuity Test The tester can measure resistance and alert the user if the resistance value is less than 200kΩ 1 Before taking a continuity test make sure that power to the de...

Страница 6: ... 6 Slide the meter back together and replace the screw You as the end user are legally bound EU Battery ordinance to return all used batteries disposal in the household garbage is prohibited You can hand over your used batteries accumulators at collection points in your community or wherever batteries accumulators are sold Disposal Follow the valid legal stipulations in respect of the disposal of ...

Страница 7: ...able input voltage 4 5V AC DC Operation time 5s 250V AC DC Power supply 2 x AAA 1 5V alkaline batteries Battery life Approx 8 hours continuous use typical Operating Temperature 10 to 55 o C 14 to 131 o F Storage Temperature 30 to 60 o C 22 to 140 o F Operating Humidity 10 to 85 RH non condensing Safety This meter is intended for origin of installation use and protected against the users by double ...

Отзывы: