background image

TM100-EU-EN v1.3   12/13 

Setup option  

The setup option is used to select the thermocouple type that matches the type that will be 
connected to the meter, to switch the sleep mode OFF/ON, and to adjust the thermocouple offset. 
Note: In the setup mode, the upper display shows the menu items and the lower display shows 
the selection for a given menu item. 
 
Press and hold the 

SET

 button for 3 seconds to enter or exit the Setup mode. When the 

thermometer is in Setup

 

mode, the display shows the 

SETUP

 display icon. 

Set Type J or K thermocouple 

1. Press 

 or 

to scroll to the “TYPE” setting (shown on the bottom left of the display). 

2. Press 

ENTER

 to choose the option. 

3. Press 

 to toggle between J and K. 

4.  Press and hold the 

SET

 button for 3 seconds to save the setting and return to normal mode. 

Set Sleep/Non-sleep mode 

The thermometer enters sleep mode by default. The meter will automatically shut off after 20 
minutes of inactivity (no buttons pressed). When the thermometer is in Setup mode, the display 
shows the SETUP display icon

.

  

1. Press 

 or 

to scroll to the 

SLP

 setting.  

2. Press 

ENTER

 to choose the option; 

On

 or 

OFF

 will appear in the display.  

3. 

Use the arrow key 

 to select ON or OFF. 

4. Press 

ENTER

 to store the new setting.  

5. 

Press and hold the 

SET

 button for 3 seconds to return to normal mode. 

Set Offset 

Use the offset option in Setup (

T1 

menu item) to adjust the thermometer’s readings to 

compensate for known temperature offset at a specific temperature. The adjustment range is ± 
5.0°C or ± 9.0 °F. 
1. 

Plug the thermocouple into the input connector. 

2. 

Place the thermocouple into the known, stable temperature environment (such as an ice 
bath or dry well calibrator). Allow the reading to stabilize. 

3. 

Use the arrow keys 

 

 

to scroll to the “T1

 option and then press 

ENTER

4. 

Change the offset temperature reading (upper display) by pressing 

until the desired 

temperature is reached. 

5. 

Press and hold the 

SET

 button for 3 seconds to return to normal mode. 

 

 

 

 

Shop for Extech products at:

1.888.610.7664

ShopExtech

.com

Содержание TM100

Страница 1: ...User s Guide Digital J K Type Thermometer Model TM100 Shop for Extech products at 1 888 610 7664 ShopExtech com ...

Страница 2: ...tech com to check for the latest version of this User Guide Product Updates and Customer Support Safety To prevent personal injury or meter damage use the meter only as specified in this guide Description Meter 1 Thermocouple input connector 2 Function buttons Power Enter C F K 3 Function buttons Hold MAX MIN Set 4 LCD display 5 Battery compartment Display 1 Setup 2 Offset option 3 Hold mode 4 Aut...

Страница 3: ...play indicates OL the temperature is outside the measurement range The display shows when a thermocouple is not connected MAX MIN AVG mode Press MAX MIN to capture and step through the maximum MAX minimum MIN and average AVG temperature values These values will be shown in the lower display with an elapsed time reference below hrs mins mins secs 1 Press MAX MIN to enter this mode 2 Press MAX MIN t...

Страница 4: ...ter 20 minutes of inactivity no buttons pressed When the thermometer is in Setup mode the display shows the SETUP display icon 1 Press or to scroll to the SLP setting 2 Press ENTER to choose the option On or OFF will appear in the display 3 Use the arrow key to select ON or OFF 4 Press ENTER to store the new setting 5 Press and hold the SET button for 3 seconds to return to normal mode Set Offset ...

Страница 5: ...ty Reminders o Never dispose of batteries in a fire Batteries may explode or leak o Never mix battery types Always install new batteries of the same type Specifications Temperature Range K type thermocouples 200 C to 1372 C 328 F to 2501 F J type thermocouples 210 C to 1100 C 346 F to 2012 F Note that thermocouple wire may be insulated with materials that cannot withstand temperatures in the range...

Отзывы: