Calibration

 

 

A

 

complete

 

calibration

 

will

 

require

 

testing

 

with

 

both

 

0

 

NTU

 

and

 

100

 

NTU

 

test

 

solutions.

 

 

Ensure

 

that

 

the

 

sides

 

of

 

the

 

bottle

 

are

 

clean

 

and

 

dry

 

to

 

prevent

 

measurement

 

errors.

 

 

Calibration

 

Procedure

 

(0

 

NTU

 

and

 

100

 

NTU)

 

Note:

 

Make

 

sure

 

you

 

complete

 

the

 

calibration

 

steps

 

with

 

as

 

little

 

time

 

as

 

possible

 

between

  

           

the

 

0

 

NTU

 

and

 

the

 

100

 

NTU

 

test

 

bottles.

 

 

1.

 

Insert

 

the

 

Zero

 

NTU

 

solution

 

bottle

 

into

 

the

 

testing

 

chamber,

 

line

 

up

 

the

 

white

 

marks,

 

and

 

close

 

and

 

latch

 

the

 

cover.

 

Turn

 

on

 

the

 

meter

 

and

 

Press

 

and

 

hold

 

the

 

"CAL"

 

button

 

until

 

the

 

LCD

 

displays

 

"CAL"

 

then

 

release

 

the

 

“CAL”

 

button.

  

 

 

 

 

2.

 

Wait

 

one

 

second

 

and

 

then

 

the

 

LCD

 

will

 

display:

 

 

 

 

 

Note:

 

If

 

the

 

LCD

 

does

 

not

 

display

 

zero

 

when

 

using

 

"0

 

NTU"

 

liquid,

 

press

 

the

 

"ZERO"

 

button

 

continuously

 

until

 

the

 

LCD

 

displays

 

a

 

zero

 

value.

 

 

3.

 

The

 

meter

 

is

 

now

 

ready

 

for

 

"0

 

NTU"

 

calibration.

 

4.

 

Momentarily

 

press

 

"CAL"

 

button.

 

The

 

LCD

 

will

 

flash

 

CAL

 

for

 

about

 

10

 

seconds

 

and

 

then

 

show

 

the

 

following

 

display:

 

 

 

 

 

5.

 

The

 

LCD

 

will

 

then

 

display:

 

 

 

 

 

6.

 

Lift

 

the

 

lid

 

on

 

the

 

test

 

chamber

 

and

 

remove

 

the

 

0

 

NTU

 

test

 

solution

 

bottle.

 

 

CAL 

 

 

     ntu 

 

0.00 

 

CAL 

     ntu 

 

CAL 

 

0.00 

    ntu 

 

100 

 

CAL 

    ntu 

www.

.com

[email protected]

1.800.561.8187

Содержание TB400

Страница 1: ...USER GUIDE Turbidity Meter Model TB400 www com information itm com 1 800 561 8187 ...

Страница 2: ...liable service Please visit our website www extech com to check for the latest version of this User Guide Product Updates and Customer Support Applications Typical applications include the measurement of municipal water food and beverage water or other aqueous solutions where fluid clarity is important Meter Description 1 LCD Display 2 HOLD button 3 REC MAX MIN button 4 POWER button 5 ZERO button ...

Страница 3: ... display will flash tESt for approximately 10 seconds and then the turbidity value in NTU units will be displayed 7 Momentarily press the OFF button to turn off the meter Data HOLD Press the HOLD button once while taking a measurement to freeze the value on the display Press HOLD again to release the Data Hold function Data Record MAX MIN Readings 1 The data record function records the MAX and Min...

Страница 4: ...d close and latch the cover Turn on the meter and Press and hold the CAL button until the LCD displays CAL then release the CAL button 2 Wait one second and then the LCD will display Note If the LCD does not display zero when using 0 NTU liquid press the ZERO button continuously until the LCD displays a zero value 3 The meter is now ready for 0 NTU calibration 4 Momentarily press CAL button The LC...

Страница 5: ...ntarily press the CAL button The LCD will flash CAL for 10 seconds 9 When calibration is complete the LCD will return to normal operating mode NOTE If the following error message appears Error cannot be calibrated check the value of the standard solution If error message still appears complete the Calibration Clear procedure CAL 100 ntu www com information itm com 1 800 561 8187 ...

Страница 6: ...on continuously until the LCD displays CAL 4 Wait one second the LCD will then display 5 Momentarily press the HOLD button The LCD will display 6 Momentarily press the HOLD button again The LCD will display 7 Momentarily Press the CAL button The LCD will flash CAL for 10 seconds 8 The calibration data clear is complete The meter will return to normal operating mode CAL ntu 0 00 CAL ntu CLr CAL ntu...

Страница 7: ...s may explode or leak Never mix battery types Always install new batteries of the same type Never dispose of used batteries or rechargeable batteries in household waste As consumers users are legally required to take used batteries to appropriate collection sites the retail store where the batteries were purchased or wherever batteries are sold Disposal Do not dispose of this instrument in househo...

Страница 8: ... 850 nm Light Detector Photo Diode Response Time Less than 10 seconds Sample Volume 10mL minimum Power Six 1 5V AAA batteries Power consumption Standby Mode approx 3 5mADC Testing Mode approx 36mADC Dimensions Weight 155 x 76 x 62mm 6 1 x 3 0 x 2 4 320g 0 70lbs Auto Power OFF Automatically switches off 10 minutes after the last key press Standard Meets ISO 7027 Copyright 2013 2016 FLIR Systems Inc...

Отзывы: