Extech Instruments SDL550 Скачать руководство пользователя страница 10

10

   

                                          

SDL550-EU-EN-V1.2-10/11

 

RS-232/USB PC Interface  

The optional 407001A software kit (software and cable) allows streaming of data to a PC via 
the RS232 Output jack. 

AC Power Adaptor 

This meter is normally powered by six (6) 1.5V ‘AA’ batteries. An optional 9V power adaptor is 
available. When the adaptor is used, the meter is permanently powered and the power button will 
be disabled. 

System Reset 

If the meter’s keys become inoperable or if the display freezes the Reset button can be used to 
reset the instrument. 

1.  Use a paper clip or similar item to momentarily press the reset button located on the lower right 

side of the instrument under the snap-off compartment cover. 

2.  After pressing the Reset button, switch the instrument ON by pressing and holding the POWER 

key for at least 1.5 seconds. If using the power adaptor unplug the adaptor and then plug it back 
in again to power the meter. 

 

Battery Replacement and Disposal 

When the low battery icon

 appears on the LCD, the batteries must be replaced. Several hours of 

accurate readings are still possible in this condition; however batteries should be replaced as soon 
as possible: 

1.  Remove the two (2) Phillips screws from the rear of the meter (directly above the top of the tilt 

stand). 

2.  Remove and safely place the battery compartment and screws where they will not be damaged 

or lost. 

3.  Replace the six (6) 1.5V ‘AA’ batteries observing polarity. 

4.  Replace the battery compartment cover with the two (2) Phillips screws. 

  

All EU users are legally bound by the battery ordinance to return all used batteries to 
collection points in your community or wherever batteries / accumulators are sold! 
Disposal in the household garbage is prohibited! 

 

Содержание SDL550

Страница 1: ...User s Guide Moisture Content Meter Plus Dew Point Wet Bulb and Thermocouple Temperature functions SD Card real time datalogger Model SDL550 RH Moisture Content Meter SDL550 ...

Страница 2: ...national Safety Symbols This symbol adjacent to another symbol or terminal indicates the user must refer to the manual for further information Meter Description 1 Thermocouple Temperature Probe input jack 2 Moisture Sensor input jack 3 Primary LCD Display digits 4 Unit of Measure for Primary Display 5 HOLD and Backlight key 6 MAX MIN key 7 SET and Clock key 8 PC interface jack 9 Reset button 10 Po...

Страница 3: ...ng the backlight ON or OFF unless the beeper is disabled Moisture Content Sensor The supplied Moisture Content sensor connects to the meter via the DIN jack at the top right of the meter Note that the utility end of the sensor is located beneath the protective cap use great caution when removing the cap and when handling the sensor as the sensor head is extremely sharp The sensor uses an internal ...

Страница 4: ...terial under test far enough to cover the sensor s openings 6 The meter will now display moisture content in RH on the primary upper LCD digits 7 The secondary lower LCD digits will show the temperature measured by the moisture probe s built in temperature sensor Use the Setup mode to change the temperature units of measure 8 Note that the probe uses a built in filter When the filter needs replaci...

Страница 5: ...asure Data Hold To freeze a displayed reading on the LCD momentarily press the HOLD button the HOLD icon will appear above the reading To exit the Data Hold mode press the HOLD button again Max Min Reading Record For a given measurement session this meter can record the highest MAX and the lowest MIN readings for later recall 1 Press the MAX MIN button momentarily to access this mode of operation ...

Страница 6: ...omatically switches out of the Setup mode if no key is pressed within 7 seconds 5 The available Setup parameters are listed below Additional detailed information is provided below this list dAtE Set the clock Year Month Date Hours Minutes Seconds SP t Set the datalogger sampling rate 0 to 3600 Seconds PoFF Automatic power off management Enable or disable the auto power off function bEEP Set the be...

Страница 7: ...umerical Format comma or decimal European and USA numerical formats differ The meter defaults to USA mode where a decimal point is used to separate units from tenths i e 20 00 The European format uses a comma i e 20 00 to separate units from tenths To change this setting 1 Access the dEC parameter 2 Use the arrow buttons to select USA or EUro 3 Press ENTER to confirm setting 4 Press and hold the S...

Страница 8: ... recommended that the card be formatted and the logger s clock set to allow for accurate date time stamping during datalogging sessions Refer to the Setup Mode section for SD card formatting and time date setting instructions European and USA numerical formats differ The data on the SD card can be formatted for either format The meter defaults to USA mode where a decimal point is used to separate ...

Страница 9: ...Up to 99 spreadsheet documents each with 30 000 readings can be stored in this folder 7 When datalogging begins a new spreadsheet document named HTB01001 xls is created on the SD card in the HTB01 folder The data recorded will be placed in the HTB01001 xls document until 30 000 readings are reached 8 If the measurement session exceeds 30 000 readings a new document will be created HTB01002 xls whe...

Страница 10: ...ing and holding the POWER key for at least 1 5 seconds If using the power adaptor unplug the adaptor and then plug it back in again to power the meter Battery Replacement and Disposal When the low battery icon appears on the LCD the batteries must be replaced Several hours of accurate readings are still possible in this condition however batteries should be replaced as soon as possible 1 Remove th...

Страница 11: ...Display update rate Approx 1 second Data Output RS 232 USB PC computer interface Operating Temperature 0 to 50 C 32 to 122 F Operating Humidity 85 R H max Auto Power OFF After 10 minutes of inactivity can be disabled Power Supply Six 6 1 5 VDC batteries optional 9V AC adaptor Power Consumption Normal operation backlight datalogger OFF approx 3 5mAdc With backlight OFF and datalogging ON approx 28m...

Страница 12: ...nds up to 1300 C 2372 F however thermocouple sensor ranges vary greatly be sure to select a thermocouple rated for the expected temperature measurement ranges Accuracy specified for meter only Thermocouple probes add additional measurement error The above specifications are tested under an environmental RF Field Strength lower than 3 V M and a frequency lower than 30 MHz Copyright 2011 Extech Inst...

Отзывы: