![Extech Instruments SDL150 Скачать руководство пользователя страница 3](http://html.mh-extra.com/html/extech-instruments/sdl150/sdl150_user-manual_2453443003.webp)
3
SDL150-EU-EN-V2.1-11/11
1
2
3
4
5
Probe Description
1.
Probe to meter input plug
2.
Probe body handle
3.
Temperature Sensor
4.
Probe head
5.
Protective cap
Getting Started
Accessories
•
The SDL150 includes batteries, SD card, probe (with spare probe head and diaphragm),
electrolyte, and carry case. If items are missing please contact the distributor through which this
product was purchased
•
Optional AC adaptor and universal AC adaptors (UK, EU, US), as well as replacement probes,
membranes, and electrolyte are available through Extech distributors
Power ON-OFF
•
Press and hold the power button
for at least 1.5 seconds to power ON the meter.
•
Press and hold the power button for 1.5 seconds to power OFF the meter.
•
This meter is powered by six (6) 1.5VDC ‘AA’ batteries or by optional AC adaptor. If the meter
will not switch ON please check that fresh batteries are installed in the rear battery compartment
or, in the case of the AC adaptor, check that the adaptor is connected correctly to the meter and
to an AC source.
Measurements
Connect the Oxygen Probe
The supplied probe connects to the keyed input jack (DIN style) at the top of the meter.
Measurement Ranges
Dissolved Oxygen measurements are displayed in the range of 0 to 20mg/L in the upper, larger digit
display area of the meter’s LCD. Measurements outside of this range will yield dashes (- - -) on the
meter’s LCD. Oxygen in Air measurements (O2) are displayed in the range of 0 to 100%.
Temperature measurements are displayed in the range of 0 to 50
o
C (32 to 122
o
F) on the lower LCD
line in smaller sized digits.
Calibration Considerations
For best accuracy, calibrate the meter with the probe before each use. The calibration should at be
performed before first use and after long periods of non-use. The calibration procedure is provided
in a later section of this user guide.