Set the Altitude Compensation (in meters) for DO measurements only
1.
Access the Height parameter.
2.
Use the arrow keys to select the altitude compensation (0 to 8900 meters in 100 meter steps)
3.
Press ENTER to confirm setting.
4.
Press and hold the SET key for at least 1.5 seconds to exit to the normal operation mode (or
simply wait 7 seconds for the meter to automatically switch to the normal operating mode).
Set the Altitude (F) Compensation (in feet) for DO measurements only
1.
Access the Height-F parameter.
2.
Use the arrow keys to select the altitude compensation (0 to 29,300 feet in 100 feet steps)
3.
Press ENTER to confirm setting.
4.
Press and hold the SET key for at least 1.5 seconds to exit to the normal operation mode (or
simply wait 7 seconds for the meter to automatically switch to the normal operating mode).
System Reset
If the meter’s keys become inoperable or if the display freezes the Reset button can be used to
reset the instrument.
1.
Use a paper clip or similar item to momentarily press the reset button located on the lower right
side of the instrument under the snap-off compartment cover.
2.
After pressing the Reset button, switch the instrument ON by pressing and holding the
POWER key for at least 1.5 seconds. If using the power adaptor unplug the adaptor and then
plug it back in again to power the meter.
Datalogging
Types of Data Recording
•
Manual Datalogging: Manually log up to 99 readings onto an SD card via push-button press.
•
Automatic Datalogging: Automatically log data onto an SD memory card where the number of
data points is virtually limited only by the card size. Readings are logged at a rate specified by
the user.
SD Card Information
•
Insert an SD card (from 1G size up to 16G) into the SD card slot at the bottom of the meter. The
card must be inserted with the front of the card (label side) facing toward the rear of the meter.
•
If the SD card is being used for the first time it is recommended that the card be formatted and
the logger’s clock set to allow for accurate date/time stamping during datalogging sessions.
Refer to the Setup Mode section for SD card formatting and time/date setting instructions.
•
European and USA numerical formats differ. The data on the SD card can be formatted for
either format. The meter defaults to USA mode where a decimal point is used to separate units
from tenths, i.e. 20.00. The European format uses a comma, i.e. 20,00. To change this setting,
refer to the Setup Mode section.