Page 29 PRC70-EN-V2.1 12/11
14 Specifications
General Specifications for
Measurement Mode
; these specifications assume:
•
A 1-year calibration cycle
•
An operating temperature of 18
°
C to 28
°
C
•
Relative humidity of 35% to 70% (non condensing)
Accuracy is expressed as ± (percentage of r percentage of range)
Function Reference Range
Resolution Accuracy
Remark
50mV
-5.000mV
~
55.000mV
1
μ
V
0.02 rdg + 0.02 range
500mV
-50.00mV
~
550.00mV
10
μ
V
0.02 rdg +0.01 range
Input Resistance: 100M
Ω
5V
-0.5000V
~
5.5000V
0.1mV
0.02 rdg +0.01 range
DCV
50V
-5.000V
~
55.000V
1mV
0.03 rdg +0.01 range
Input Resistance: 1M
Ω
DCmA
50mA
-5.000mA
~
55.000mA
1
μ
A
0.02 rdg +0.01 range
Shunt Resistance: 10
Ω
500
Ω
Test Current:
Approx. 1mA
0.00
Ω
~
550.00
Ω
0.01
Ω
0.05 rdg +0.02 range
OHM
5K
Ω
Test Current:
Approx. 0.1mA
0.0000 K
Ω
~
5.5000K
Ω
0.1
Ω
0.05 rdg +0.02 range
Open Circuit Voltage: 2.5V approx.
Does not include lead resistance
500Hz
3Hz
~
500.00Hz
0.01Hz
5KHz
3Hz
~
5.0000KHz
0.1Hz
FREQ
50KHz
3Hz
~
50.000KHz
1Hz
±2 digits
Input Impedance
:
100k
Ω
min.
Sensitivity
:
3Vp-p min.
Duty Cycle: 50%
R
32
°
F to 3212
°
F
0
°
C to 1767
°
C
S
32
°
F to 3212
°
F
0
°
C to 1767
°
C
1.8
°
F (1
°
C)
32
°
F to 932
°
F: 3.2
°
F
932
°
F to 3212
°
F: 2.7
°
F
0 to 500
°
C: 1.8
°
C
500 to 1767
°
C: 1.5
°
C
K
-148
°
F to 2501.6
°
F
-100.0
°
C to 1372.0
°
C
-148.0
°
F to 32.0
°
F: 2.2
°
F
32.0
°
F to 2501.0
°
F: 1.4
°
F
-100.0
°
C to 0.0
°
C: 1.2
°
C
0.0
°
C to 1372.0
°
C: 0.8
°
C
E
-58.0
°
F to 1832.0
°
F
-50.0
°
C to 1000.0
°
C
-58.0
°
F to 32.0
°
F: 1.6
°
F
32.0
°
F to 1832.0
°
F: 2.7
°
F
-50.0
°
C to 0.0
°
C: 0.9
°
C
0.0
°
C to 1000.0
°
C: 1.5
°
C
J
-76.0
°
F to 2192.0
°
F
-60.0
°
C to 1200.0
°
C
-76.0
°
F to 32.0
°
F: 1.8
°
F
32.0
°
F to 2192.0
°
F: 1.3
°
F
-60.0
°
C to 0.0
°
C: 1.0
°
C
0.0
°
C to 1200.0
°
C: 0.7
°
C
T
-148.0
°
F to 752.0
°
F
-100.0
°
C to 400.0
°
C
-148.0
°
F to 32.0
°
F: 1.8
°
F
32.0
°
F to 752.0
°
F: 1.3
°
F
-100.0
°
C to 0.0
°
C: 1.0
°
C
0.0
°
C to 400.0
°
C: 0.7
°
C
N
-328.0
°
F to 2372.0
°
F
-200.0
°
to 1300.0
°
C
0.1
°
F (0.1
°
C)
-328.0
°
F to 32.0
°
F: 2.7
°
F
32.0
°
F to 2372.0
°
F: 1.6
°
F
-200.0
°
C to 0.0
°
C: 1.5
°
C
0.0
°
C to 1300.0
°
C: 0.9
°
C
TC
B
1112
°
F to 3308
°
F
600
°
C to 1820
°
C
1
°
F (1
°
C)
1112
°
F to 1472
°
F: 4.0
°
F
1472
°
F to 1832
°
F: 3.2
°
F
1832
°
F to 3308
°
F: 2.5
°
F
600
°
C to 800
°
C: 2.2
°
C
800
°
C to 1000
°
C: 1.8
°
C
1000
°
C to 1820
°
C: 1.4
°
C
ITS-90 temperature scale;
Accuracy does not include the error
of the internal temperature
compensation caused by the sensor
Содержание PRC70
Страница 1: ...User Guide Multifunction Process Calibrator Model PRC70 ...
Страница 11: ...Page 11 PRC70 EN V2 1 12 11 ...
Страница 12: ...Page 12 PRC70 EN V2 1 12 11 ...
Страница 14: ...Page 14 PRC70 EN V2 1 12 11 ...
Страница 15: ...Page 15 PRC70 EN V2 1 12 11 ...
Страница 20: ...Page 20 PRC70 EN V2 1 12 11 ...
Страница 21: ...Page 21 PRC70 EN V2 1 12 11 ...
Страница 23: ...Page 23 PRC70 EN V2 1 12 11 ...