![Extech Instruments PRC50 Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/extech-instruments/prc50/prc50_user-manual_2453746024.webp)
Page 24 PRC50-EN-V2.1 11/11
General Specifications for Source Mode
These specifications assume:
•
1-year calibration cycle
•
Operating temperature of 64
o
F to 82
o
F (18
°
C
to 28
°
C
)
•
Relative humidity of 35% to 70% (non condensing)
Accuracy is expressed as ± (percentage of set value + percentage of range)
Function
Reference Range
Resolution
Accuracy
Remark
1000mV
-100.00mV
~
1100.00mV
10
μ
V
0.02 rdg +0.01 range
Maximum output current: 2mA
DC voltage
10V
-1.0000V
~
11.0000V
0.1mV
0.02 rdg +0.01 range
Maximum output current: 5mA
DC current
20mA
0.000mA
~
22.000mA
1
μ
A
0.02 rdg +0.02 range
External supply for simulate mA: 5V–
28V
Maximum load 1K
Ω
at 20mA
100Hz
1.00Hz
~
110.00Hz
0.01Hz
1KHz
0.100KHz
~
1.100KHz
1Hz
FREQ
10KHz
1.0KHz
~
11.0KHz
0.1KHz
±2 counts
100Hz
1KHz
PULSE
10KHz
1 to 100,000cycles
1cyc
Output voltage: +1~+11 V
p-p
(zero
based waveform);
Amplitude accuracy: ±(10% +0.5V);
Maximum load: >100 K
Ω
;
Duty Cycle: 50%.
LOOP
24V
±
10%
Maximum current: 25 mA
Short circuit protected
Notes:
z
Temperature Coefficient: 0.1 times the applicable accuracy specification per degree from 41
o
F to 64
o
F (5
°
C
to 18
°
C)
and 82
o
F to
104
o
F (28
°
C
to 40
°
C)
.
z
Maximum voltage between any output terminal and earth ground: 30Vpeak
z
Maximum output current: Approximately 25mA.
Copyright
©
2011
Extech
Instruments
Corporation
(a
FLIR
company)
All
rights
reserved
including
the
right
of
reproduction
in
whole