![Extech Instruments PRC50 Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/extech-instruments/prc50/prc50_user-manual_2453746023.webp)
Page 23 PRC50-EN-V2.1 11/11
10 Specifications
General Specifications for
Measurement Mode
; these specifications assume:
•
A 1-year calibration cycle
•
An operating temperature of 64
o
F to 82
o
F (18
°
C
to 28
°
C)
•
Relative humidity of 35% to 70% (non condensing)
Accuracy is expressed as ± (percentage of r percentage of range)
Function
Reference
Range Resolution
Accuracy
Remark
200mV
-20.00mV
~
220.00mV
10
μ
V
0.02 rdg + 0.02 range
Input Resistance: 100M
Ω
5V
-0.5000V
~
5.5000V
0.1mV
0.02 rdg +0.01 range
DCV
50V
-5.000V
~
55.000V
1mV
0.03 rdg +0.01 range
Input Resistance: 1M
Ω
DCmA
50mA
-4.000mA
~
55.000mA
1
μ
A
0.02 rdg +0.01 range
Input Resistance: 5
Ω
500Hz
3Hz
~
500.00Hz
0.01Hz
5KHz
3Hz
~
5.0000KHz
0.1Hz
FREQUENCY
50KHz
3Hz
~
50.000KHz
1Hz
±2 digits
Input Impedance
:
100k
Ω
min.
Sensitivity
:
3Vp-p min.
Duty Cycle: 50%
CONTINUITY
≤
250
Ω
beeper
Open circuit voltage: 2.5V
Notes:
z
DCV: Normal Mode Rejection Ratio (NMRR)
≥
60dB (at 50Hz or 60Hz); Common Mode Rejection Ratio (CMRR)
≥
120dB (at
50Hz or 60Hz)
z
Reading rate: 2 readings per second
z
Temperature Coefficient: 0.1 times the applicable accuracy specification per degree from 41
o
F to 64
o
F (5
°
C
to 18
°
C)
and 82
o
F to
104
o
F (28
°
C
to 40
°
C)
.
z
Maximum voltage between V
Ω
Hz terminal and COM terminal: 60 Vp-p; Maximum Input current: 55mA.
Protected with a 63mA, 250V fast blow fuse