PQ2071-en-GB_V1.3 7/15
22
General Specifications
Clamp jaw opening
57mm (2-1/4") approx.
Display
Dual 4-digit (9999 counts) backlit LCD
Bargraph
100 units
Low Battery indication
Battery symbol is displayed
Over-range indication
‘OL’
display
Internal Reading Memory
99 readings can be stored, recalled, and cleared.
External PC Memory
50,000 readings can be recorded on a connected PC running the
supplied PC software. Readings can be exported to a spreadsheet.
Measurement rate
2 readings per second, nominal
Input Impedance
10M
(VAC)
AC bandwidth
50 to 200Hz (AAC and VAC)
AC response
True rms (AAC and VAC)
Operating Temperature
5
C to 40
C (41
F to 104 F)
Storage Temperature
-20
C to 60
C (-4
F to 140
F)
Operating Humidity
< 80% up to 31
C (87
F) decreasing linearly to 50% at 40
C (104
F)
Storage Humidity
< 80%
Operating Altitude
2000meters (7000ft) maximum
Battery
One (1) 9V Battery (NEDA 1604)
Auto power OFF
After approx. 25 minutes
Dimensions & Weight
292 x 95 x 38mm (11.5 x 3.75 x 1.5”); 522g (18.4 oz)
Safety
For indoor use and in accordance with the requirements for double
insulation to IEC1010-1 (2001): EN61010-1 (2001) Overvoltage
Category IV 600V and Category III 1000V, Pollution Degree 2