Chapter 6
-
Troubleshooting
6.1 Problem Solving
1. If the oscilloscope will not power ON, follow these steps:
1 Check the power cord to verify it has been connected properly;
2 Check the power on/off button to ensure it has been pushed;
3 Restart the oscilloscope.
Contact your local Extech distributor or the Extech Technical Support department if the problem persists.
2. If waveforms will not display when the oscilloscope is switched ON:
1 Check the probe to assure it is properly connected to the input BNC;
2 Check the channel switch (such as CH1, CH2 menu buttons) to make sure it has been
turned on;
3 Check the input signal to verify it has been connected to the probe correctly;
4 Verify that all measured circuits indeed have signals to output;
5 Raise the magnitude for DC signals;
6 Press the
Auto
button to perform an automatic detection of signals.
Contact Extech Technical Support department if there is still no display of waveforms.
3. If the waveform of the input signal is seriously distorted, follow these steps:
1 Check the probe to assure a proper connection to the channel BNC;
2 Check the probe to assure a good connection to the measured object;
3 Check the probe to verify that it has been well calibrated. Otherwise, refer to the content
regarding Calibration in this manual.
4. If the waveform is moving continuously on the screen and can not be triggered:
1 Check the trigger source to ensure that it is consistent with the input channel;
2 Check the trigger level to assure a correct setting.
3 Check the trigger mode to confirm that it is an appropriate selection for the input signal.
The default trigger mode is Edge trigger. However, edge triggering it is not suitable for
every type of input signal.
5. If the meter will not display current measurements:
The internal fuses may have opened. The meter is protected by two fast acting fuses not
accessible to the user. The fuses are rated: 1A (600V) and 12A (600V) fast acting.
Please contact a local Extech distributor or the Extech Technical Support department for
assistance in locating a local service center.
www.burntec.com
Содержание MS6060
Страница 1: ...User Manual MS6000 Series Digital Storage Oscilloscope www burntec com ...
Страница 36: ...36 MS6000 EU EN V1 3 7 12 www burntec com ...
Страница 54: ...54 MS6000 EU EN V1 3 7 12 Pass Fail test Mask Blue and signal yellow display www burntec com ...
Страница 82: ...82 MS6000 EU EN V1 3 7 12 www burntec com ...
Страница 84: ...84 MS6000 EU EN V1 3 7 12 www burntec com ...