11

Resistance

 

Measurements

 

WARNING:

 

Please remove power to the circuit under test and discharge all capacitors 

before taking resistance or continuity measurements. The meter will display 

OL

 if the circuit 

under test is an open circuit of if the measurement exceeds the maximum range of the meter. 
Do not input a voltage >30V AC or DC.

 

1.

 

Turn the function switch to the Resistance 

 position. 

2.

 

Use the 

M

 (mode) button to select the ohm 

 display symbol. 

3.

 

Insert the black test lead banana plug into the negative (COM) jack. Insert the red test 
lead banana plug into the positive (

) jack. See Fig. 7. 

4.

 

Touch the test probe tips across the circuit or part under test. It is best to disconnect one 
side of the part under test so that the rest of the circuit will not interfere with the 
resistance reading. 

5.

 

Read the resistance in the display. The display will indicate the proper decimal point and 
value. 

6.

 

The MAX

MIN function is available in this mode on the MA443 and MA445. Short press 

the 

MAX

 

MIN

 button to step through MAX MIN readings. Long press to exit the mode. 

NOTES:

 

For low resistance measurements, the probe error (0.1~0.2

) should be zeroed using the 

relative mode. Short circuit the test probes and press the REL button before taking 
measurements. 
When measuring resistance >1M

 the meter may require several seconds to stability the 

displayed reading.  

MAX-MIN

For MA443 and
MA445

 

Fig.

 

7

 – Resistance and Continuity Measurements

 

Содержание MA440

Страница 1: ...USER MANUAL Model MA440 400A AC Clamp Meter DMM Model MA443 True RMS 400 AC Clamp Meter DMM Model MA445 True RMS 400A AC DC Clamp Meter DMM...

Страница 2: ...ed and calibrated and with proper use will provide years of reliable service Please visit our website www extech com to check for the latest version of this User Guide Product Updates Product Registra...

Страница 3: ...60 days 9 Do not service or modify the internal circuits of the meter Qualified personnel must service the meter only 10 Do not use the meter in flammable or explosive environments gas or vapors Do n...

Страница 4: ...e or high humidity Safety Symbols This symbol adjacent to another symbol indicates the user must refer to the manual for further information Risk of electrical shock Alert beeper Equipment protected b...

Страница 5: ...ht button MA443 and MA445 11 NCV sensor Note Battery compartment is located on back of meter Fig 1 Meter Description Display Description Relative mode icon Units of Frequency Alternating Direct Curren...

Страница 6: ...ent measurements will be displayed as measurement minus reference value This feature applies to AC DC voltage AC current resistance and temperature modes only Short press to exit For DC current measur...

Страница 7: ...r switches OFF after 30 minutes MA440 or 15 minutes MA443 MA445 of inactivity The MA443 MA445 models emit a warning tone prior to automatically switching off Disable Auto Power OFF MA443 and MA445 To...

Страница 8: ...ing AC Amperes Amps The display icon AUTO will also appear indicating automatic ranging 3 Press the jaw trigger to open the clamp jaw 4 Position the clamp around only one conductor See Fig 2 for corre...

Страница 9: ...3 and Fig 4 4 Press the jaw trigger to open the clamp jaw Position the clamp around only one conductor See Fig 2 previous section for correct and incorrect clamping techniques 5 Read the current in t...

Страница 10: ...oint and value 6 Relative REL mode can be used to set a reference reading from which subsequent readings will be offset reference reading actual reading displayed reading Long press the REL button to...

Страница 11: ...test probe tips across the circuit or part under test It is best to disconnect one side of the part under test so that the rest of the circuit will not interfere with the resistance reading 5 Read th...

Страница 12: ...k See Fig 7 above 4 Touch the test probe tips across the circuit or wire 5 If the resistance is 30 the meter will beep If the resistance is 60 the meter will not beep If the resistance is between 30 a...

Страница 13: ...ive COM jack Insert the red test lead banana plug into the positive jack See Fig 9 4 Touch the test probe tips across the circuit or component 5 Read the Capacitance measurement on the meter display F...

Страница 14: ...nsert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the positive V jack See Fig 10 4 Touch the test probes to the diode under test in both polarity...

Страница 15: ...ocouple Do not measure temperature if the meter is in an environment outside the range of 18 28o C Do not input voltages 30V DC or AC to the meter 1 Turn the function switch to the TEMP position 2 Use...

Страница 16: ...g When the sensed field is at highest levels the meter s NCV lamp and beeper rate will become as one continuous light and tone If the meter does not react as described above to an AC voltage or electr...

Страница 17: ...rtment 3 Replace the three 3 1 5V AAA batteries observing correct polarity 4 Re assemble the meter before use 5 Safety Please dispose of batteries responsibly never dispose of batteries in a fire batt...

Страница 18: ...range Frequency response 50 60Hz MA440 and MA443 and 40 400Hz MA445 MA443 MA445 AC Crest factor considerations a Add 4 MA443 3 MA445 when wave factor is 1 2 b Add 6 MA443 5 MA445 when wave factor is 2...

Страница 19: ...A440 50 00nF 0 01 nF 4 0 25 digits 600V AC DC 500 0nF 0 1 nF 4 0 5 digits 5 000 F 0 001 F 50 00 F 0 01 F 100 0 F 0 1 F Capacitance MA443 and MA445 40 00nF 0 01 nF 4 0 25 digits 600V AC DC 400 0nF 0 1...

Страница 20: ...0 400 o C 400 1000 o C 1 0 3 digits 40 104 o F 1o F 3 0 10 digits 104 752 o F 752 1832 o F 1 0 6 digits Does not include accuracy of the Type K temperature probe Non Contact Voltage Detector NCV 100Vr...

Страница 21: ...mum applied to any terminal Low battery indication is displayed Auto Power OFF After 30 minutes MA440 or 15 minutes MA443 and MA445 Hold M button while turning meter ON to disable APO MA443 and MA445...

Отзывы: