AC Current Measurements 

CAUTION: Observe CAT II 600V, CAT III 300V with respect to Earth 

Ground for the Jaw. 

1.

 

Set the Function switch to the 2A, 20A, or 200A 

range. 

2.

 

Press the Trigger to open the clamp jaw. 

3.

 

Clamp onto a single conductor (fully enclosing it). See 

diagram below for proper clamping technique. 

4.

 

Use the indicator mark to determine the geometric 

center of the clamp head (see Meter Description 
section). Position the measured electrical conductor 
as close to this geometric center as possible. 

5.

 

Do not allow a gap between the two halves of the 

jaw.  

6.

 

Read the ACA value on the LCD. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Data Hold 

To freeze the displayed reading on the LCD, long press the 
Data Hold (H) button (the 

H

 icon will appear on the 

display). To release the Data Hold function and return the 
meter to normal operation, long press the Data Hold (H) 
button again (the 

H

 icon will switch off). 

MAX Reading 

1.

 

Press MAX to show the highest reading on the 
display. The MAXH display icon will appear and the 
reading shown will be the highest reading; updating 
only when a higher reading is encountered. 

2.

 

Press MAX again to exit this mode and to view real 
time readings. The MAXH display icon will switch 
off. 

Содержание MA130

Страница 1: ...USER MANUAL 200A AC Mini Clamp Meter Model MA130 Please visit www extech com for user manual translations...

Страница 2: ...w battery indicator and Auto Power Off feature CAT II 600V and CAT III 300V safety rating Safety Information To ensure the safe operation and service of the meter follow these instructions closely Fai...

Страница 3: ...d actions that could cause DAMAGE to the meter or equipment under test Do not expose the meter to extremes in temperature or high humidity Do not expose the meter to extremes in temperature or to high...

Страница 4: ...nces OVERVOLTAGE CATEGORY III Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations Note Examples include switches in the fixed installation and some equipment for industrial use w...

Страница 5: ...d in the safety section of this manual prior to use Powering the Meter 1 Turn the rotary function switch to any position to power the meter Check the batteries if the unit fails to power ON Refer to t...

Страница 6: ...s geometric center as possible 5 Do not allow a gap between the two halves of the jaw 6 Read the ACA value on the LCD Data Hold To freeze the displayed reading on the LCD long press the Data Hold H bu...

Страница 7: ...esponsibly never dispose of batteries in a fire batteries may explode or leak If the meter is not to be used for 60 days or more remove the battery and store separately Never dispose of used batteries...

Страница 8: ...d Battery Two 1 5V AAA batteries Low battery display Operating Temperature 0 C to 40 C 32 F to 104 F Operating Humidity 0 C to 30 C 32 F to 86 F 75 RH 30 C to 40 C 86 F to 104 F 50 RH Storage Temperat...

Страница 9: ...00 A 10mA 3 0 rdg 20 digits 200A AC 200 0 A 100mA 2 5 rdg 20 digits 200A AC Notes Frequency range 50 60Hz Maximum overload protection for clamp head terminal is 200A Accuracy stated for 10 100 of ran...

Отзывы: