Auto-Power OFF Feature
The auto power off feature automatically shuts the meter off 12 minutes after the most recent button
press.
Disabling the Auto-Power OFF Feature
With the unit ON, momentarily press the
CAL/RECALL
button, then quickly press and hold both the
MODE/HOLD
and
ON/OFF
buttons until ‘oFF’ is displayed. To restore the Auto Power Off Feature
(auto power OFF enable) simply turn the meter off and on again using the
ON/OFF
button.
Low Battery Indication
When the battery voltage falls below the operating threshold, ‘BAT’ will appear on the display. Refer
to the Maintenance section for battery replacement information.
Storing Readings
Up to 25 readings can be stored in memory for later recall.
1. With the meter in the HOLD mode, press and hold the
MODE/HOLD
button for three (3)
seconds to store a reading. Release the button when the memory location number appears on
the lower display.
2. After approx. 30 seconds (measurement duration) the meter will return to the HOLD mode and
another reading can then be stored.
3. If more than 25 readings are stored, previously stored readings (starting with reading number
1) are overwritten.
Recalling Stored Readings
1. Momentarily press the
CAL/RECALL
button and then within 4 seconds momentarily press the
MODE/HOLD
. The last stored data point location will be displayed (1 to 25). Each time the
MODE/HOLD
button is momentarily pressed the next most recently stored data point will be
displayed.
2. After the last data point is displayed, pressing the
MODE/HOLD
button again returns the
display to the beginning of the list.
3. Pressing
the
CAL/RECALL
button at anytime stops the data retrieval process and returns the
meter to the normal measurement mode.
Clearing Stored Readings
1. With the unit ON press and hold the
ON/OFF
button for 4 seconds
2. When “CLR” is shown in the main display the memory is cleared.
Содержание ExStik FL700