Introduction 

The model FL700 is a system specifically designed for the quick and accurate measurement of 
fluoride ions in drinking water and other aqueous samples. Unlike other electrode based systems 
the FL700 consists of the sensing electrode, measuring electronics, and the display in one 
convenient package. This meter is shipped fully tested; with proper use, this instrument will 
provide years of reliable service. 
  
 
 

Features 

 

Automatic temperature compensation (±10

o

C of calibration temperature) 

 Automatic 

calibration 

 

Stability sensing to optimize accuracy 

 

Internal Datalogger for storing up to 25 readings 

 

Direct reading of ppm, mV, or mg/l

 

units 

 

Automatic power off after 12 minutes to preserve battery life 

 

Internal error detection  

 
 

Содержание ExStik FL700

Страница 1: ...User Guide ExStik FL700 Fluoride Meter ...

Страница 2: ...ng electronics and the display in one convenient package This meter is shipped fully tested with proper use this instrument will provide years of reliable service Features Automatic temperature compensation 10 o C of calibration temperature Automatic calibration Stability sensing to optimize accuracy Internal Datalogger for storing up to 25 readings Direct reading of ppm mV or mg l units Automatic...

Страница 3: ...ws the users to follow the ASTM and EPA standard methodology using any of the TISAB reagents and standards already in use Also Extech has developed tablet form TISAB which contains all of the essential and approved chemicals that are found in the usual TISAB reagents A total ionic strength adjustment buffer TISAB is a buffer solution which increases the ionic strength of a solution to a relatively...

Страница 4: ... the 1ppm Fluoride standard into the sample cup Note if your 1ppm Fluoride standard already contains TISAB do not use a TISAB tablet 2 Create a rinse solution that can be used between sample measurements buy dissolving 1 TISAB tablet in 20ml of Distilled water The rinse solution promotes faster response times 3 Rinse the end of the FL700 module in this Rinse solution or in Distilled water and then...

Страница 5: ...centration and will then enter the HOLD mode Note The readings can be stored in the memory by pressing the MODE HOLD key for 3 seconds Electrode Storage 1 It is recommended that the electrode be stored WET in the last test solution used by the instrument fluoride ion plus TISAB reagent 2 The module and can be stored dry If stored dry it will be necessary to allow approximately 15 minutes of soakin...

Страница 6: ...ng Release the button when the memory location number appears on the lower display 2 After approx 30 seconds measurement duration the meter will return to the HOLD mode and another reading can then be stored 3 If more than 25 readings are stored previously stored readings starting with reading number 1 are overwritten Recalling Stored Readings 1 Momentarily press the CAL RECALL button and then wit...

Страница 7: ... from side to side pulling it away from the meter until it disconnects 3 To attach an electrode align the positioning keys on the electrode and the main body housing and then carefully push the electrode into the meter socket until it is fully seated 4 Tighten the electrode retaining collar firmly enough to seal the electrode with the meter Electrode Storage 3 It is recommended that the electrode ...

Страница 8: ...months minimum Measurement method In compliance with EPA Method 340 2 Potentiometric Ion Selective Electrode Response Time 90 of change in less than 30 seconds typical Operating Temp Range 32 to 140 F 0 to 60 C ATC Range 32 to 140 F 0 to 60 C Measurement Storage 25 tagged numbered data sets with recall Battery Power Four 4 CR2032 button batteries Low Battery Indication BAT appears on the LCD Auto ...

Страница 9: ...l or consequential damages FLIR s total liability is limited to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is expressed or implied Calibration Repair and Customer Care Services FLIR Systems Inc offers repair and calibration services for the Extech Instruments products we sell NIST certification for most products is a...

Страница 10: ...entel La responsabilité totale de FLIR est limitée à la réparation ou au remplacement du produit La garantie définie ci dessus est inclusive et aucune autre garantie écrite ou orale n est exprimée ou implicite Calibrage réparation et services après vente FLIR Systems Inc offre des services de calibrage et de réparation pour les produits Extech Instruments que nous commercialisons Nous fournissons ...

Страница 11: ...ncidentales o consecuentes La responsabilidad total de FLIR está limitada a la reparación o reemplazo del producto La garantía precedente es inclusiva y no hay otra garantía ya sea escrita u oral expresa o implícita Servicios de calibración reparación y atención a clientes FLIR Systems Inc ofrece servicios de reparación y calibración para los productos que vendemos de Extech Instruments Además ofr...

Отзывы: