EX35x
‐
en
‐
GB_V1.2
12/15
17
LoZ
AC
Voltage
Measurements
When
the
function
switch
is
turned
to
the
Lo
Z
position,
the
meter
applies
a
low
input
impedance
circuit
to
the
input
terminals
to
eliminate
stray
or
ghost
voltages.
Refer
to
the
Voltage
Measurements
section
earlier
in
this
guide
for
Safety
information
and
connection
diagrams.
1.
Insert
the
black
test
lead
banana
plug
into
the
negative
(COM)
jack
and
the
red
test
lead
banana
plug
into
the
positive
(V/
Ω
)
jack.
2.
Move
the
Function
Switch
to
the
Lo
‐
Z
position.
3.
Touch
the
test
probe
tips
to
the
circuit
under
test.
4.
The
meter
defaults
to
the
Auto
Range
mode
(Auto
displayed
on
LCD).
Manual
range
cannot
be
used
in
this
mode.
5.
Read
the
digital
value
of
the
measurement
in
the
display.
The
display
will
also
indicate
the
proper
decimal
point
and
measurement
type/units
symbols.
Variable
Frequency
Voltage
Measurements
When
the
function
switch
is
turned
to
the
LPF
position
the
meter
can
measure
voltage
on
signals
with
varying
frequency
as
in
Variable
Frequency
Drives
(VFD).
Refer
to
the
Voltage
Measurements
section
earlier
in
this
guide
for
Safety
information
and
connection
diagrams.
1.
Insert
the
black
test
lead
banana
plug
into
the
negative
(COM)
jack
and
the
red
test
lead
banana
plug
into
the
positive
(V/
Ω
)
jack.
2.
Move
the
Function
Switch
to
the
LPF
position.
3.
Select
AC
voltage
4.
Press
and
Hold
the
LPF
button
for
two
seconds.
The
meter
will
beep
and
the
LPF
icon
will
appear
on
the
display.
5.
Touch
the
test
probe
tips
to
the
circuit
under
test.
6.
The
meter
defaults
to
the
Auto
Range
mode.
Manual
range
cannot
be
used
in
this
mode.
7.
Read
the
digital
value
of
the
measurement
in
the
display.
The
display
will
also
indicate
the
proper
decimal
point
and
measurement
type/units
symbols.