EC600   V2.0   4/10 

Keypad Description 

 

The meter has five (5) front panel function keys: 
 

     ON/OFF Power Key 

 

   Calibration Key: 

a)  When taking a measurement, press this key to enter the calibration mode 
b)  When in the programming mode, press this key to change the displayed 

number or the ON/OFF state 

   Function Key 

a)  In pH measuring mode, one short press switches between pH and mV 

measuring modes. One long press (>2s) enters the parameter setting mode. 
Subsequent presses will scroll through the available parameters. 

b)  In other measuring modes, press and hold to enter the parameter setting 

modes. The meter will scroll through the available parameters 

   Backlight and ENTER Key 

a)  In the measurement mode, momentary presses toggle the backlight on and off 
b)  When in the programming or calibration mode, press to ENTER data 
c)  When in the pH mode, press and hold to change the resolution in the range of 

0.01 to 0.1pH 

d)  When in the Conductivity mode, press and hold to scroll through the TDS, 

Salinity, Resistivity, and Conductivity modes 

   MEMORY STORE and RECALL Key 

a)  In the measurement mode, momentary presses store the displayed readings. 

Press and hold for at least 2 seconds to recall a saved reading 

b)  In programming mode, press to change the displayed number or ON/OFF state 

 

Store, Recall, and Clear Datalogger Memory 

STORE Readings 

The meter can store up to 100 pH, 100 mV, and 100 Conductivity readings for a total of 
300 data points. To store a reading, wait until the reading stabilizes (the smiling face icon 

 appears when the reading stabilizes). Press the 

M+/RM

 key momentarily to store a 

reading. The 

M+

 icon appears and the data point serial number increments. 

RECALL Readings 

In the measurement mode, press the 

M+/RM

 key to recall the most recently stored 

reading. 

RM

 and the data point serial number for the displayed reading will appear on the 

LCD. The measurement information will appear on the lower right hand side of the LCD. 
Use the 

CAL 

or 

M+/RM

 buttons to scroll the remaining stored readings. Press the 

ENTER

 

key to return to the normal measurement mode. 

CLEAR Readings 

From the RECALL mode, press and hold the 

ENTER

 key for at least five (5) seconds. The 

LCD display will show ‘CLR’ indicating that all of the stored readings have been erased. 

www.burntec.com

www.burntec.com

Содержание EC600

Страница 1: ...User s Guide Digital Conductivity pH Meter Model EC600 www burntec com www burntec com ...

Страница 2: ...r solutions and 8 types of Conductivity standards The meter s digital filter improves measurement speed and accuracy Two special modes for distilled water and distilled water mixed with ammonia are provided The meter offers a backlit LCD display and is dust proof and water proof meeting the IP57 rating This meter is shipped fully tested and calibrated and with proper use will provide years of reli...

Страница 3: ...section below 4 Electrode connection jack 5 Electrode plug 6 Electrode cable Display Description 1 Measurement type icon 2 Measured reading 3 888 Stored data serial number M measurement to be stored icon RM reading to be recalled icon 4 Units of measure 5 Temperature reading 6 Stable measurement icon 7 Calibration icons 8 Low battery icon www burntec com www burntec com ...

Страница 4: ...E and RECALL Key a In the measurement mode momentary presses store the displayed readings Press and hold for at least 2 seconds to recall a saved reading b In programming mode press to change the displayed number or ON OFF state Store Recall and Clear Datalogger Memory STORE Readings The meter can store up to 100 pH 100 mV and 100 Conductivity readings for a total of 300 data points To store a rea...

Страница 5: ...EC600 V2 0 4 10 5 The unit will automatically return to the normal measurement mode after approximately 2 seconds www burntec com www burntec com ...

Страница 6: ...g is reached 6 Press CAL again and the display will show a blinking 4 00 7 This portion of the calibration procedure is complete when the display stops blinking and shows the C3 icon The unit will automatically switch to the third point of the calibration 8 Rinse the probe in distilled water again allow it to dry and submerge it into a pH 10 01 buffer solution Stir the solution briefly and allow i...

Страница 7: ...lection USA Europe U S A NIS NIST CH China P2 Distilled water pH temperature compensation setting see notes below this Table OFF ON P3 Distilled water with Ammonia pH temperature compensation setting see Notes below this Table OFF ON P4 Temperature unit setting P5 Back light display time setting 0 1 3 6min P6 Auto power off setting 0 10 20min P7 Restore to default factory settings OFF ON Notes on ...

Страница 8: ...r P4 or press ENTER to return to the normal measurement mode Parameter P4 Temperature Measurement Units Setting 1 From the P4 menu use the CAL or the M RM keys to turn this feature ON or OFF 2 Momentarily press MODE to move to the next parameter P5 or press ENTER to return to the normal measurement mode Parameter P5 Display Backlight Setting 1 From the P5 menu use the CAL or the M RM keys to selec...

Страница 9: ...ter and diluted to 100mL The electrode should not be soaked in a purified water protein solution or an acid fluoride solution for long periods of time In addition do not soak the electrode in organic silicon lipids For calibration accuracy the pH of the standard buffer solution must be reliable The buffer solution should be refreshed often especially after heavy use For best accuracy always keep t...

Страница 10: ...ngs P1 Back light display time 0 1 3 6 min P2 Auto power off time 0 10 20 min Parameter P1 Display Backlight Setting 1 Press MODE to access the P1 Parameter 2 Use the CAL or the M RM keys to select 0 1 3 or 6 minute default backlighting time 3 Momentarily press MODE to move to the next parameter P2 or press ENTER to return to the normal measurement mode Parameter P2 Automatic Power OFF Setting 1 F...

Страница 11: ...mplete 7 If the measurement value is unstable repeat the calibration until the measurement is stable Replace the electrode if necessary Notes The meter is calibrated before leaving the factory and can generally be used right out of the box The meter can only be calibrated in the Conductivity mode and not from the TDS Salinity or Resistivity mode Measuring the Conductivity of a Sample 1 Clean and d...

Страница 12: ...entration Use the chart below for common solution types use Parameter P4 as described below in the Programming Parameters Conductivity section The meter can perform an automatic non linear temperature compensation in distilled water for measurements below 10 μS cm Note When the coefficient for the temperature compensation is set to 0 00 no compensation the measurment value will be based on the cur...

Страница 13: ...0 20 minutes P9 Restore to default factory settings OFF ON Parameter P1 pH Buffer Solution Setting 1 From the pH measurement mode press and hold MODE for at least 2 seconds and then release the P1 icon appears on the LCD 2 Use the CAL or the M RM keys to toggle through the two 2 selections USA for use in the USA or Europe and CH for use in China 3 Momentarily press MODE to move to the next paramet...

Страница 14: ...mode Parameter P6 Temperature units of measure 1 From the P6 menu use the CAL or the M RM keys to select the desired unit of measure C or F 2 Momentarily press MODE to move to the next parameter P7 or press ENTER to return to the normal measurement mode Parameter P7 Display Backlight Setting 1 From the P7 menu use the CAL or the M RM keys to select 0 1 3 or 6 minute default backlighting time 2 Mom...

Страница 15: ...odes with distilled water and air dry The surface of the supplied conductivity electrode is plated with a layer of metal platinum black in order to lower the electrode polarization and increase the measuring range Do not polish the black platinum surface clean it by stirring in Distilled water If excessive organic buildup appears on the black platinum coating clean with hot water and detergent or ...

Страница 16: ...mV Accuracy Meter 0 1 FS Conductivity Specifications Conductivity measuring ranges 0 00 to 19 99 µS cm 20 0 to 199 9 µS cm 200 to 1999 µS cm 2 00 19 99 mS cm 20 0 to 199 9 mS cm TDS measurement range 0 to 100 g L Total Dissolved Solids Salinity measurement range 0 to 100 ppt Resistivity 0 to 100 Mohms Resolution 0 01 0 1 1 µS cm and 0 01 0 1mS cm Accuracy Meter only 1 F S with probe 2 F S Electrod...

Страница 17: ... Two AA batteries 1 5V x 2 Size and weight Meter 2 6 x 4 7 x 1 2 65 120 31mm 6 3 oz 180g Case 14 1 x 10 6 x 3 360 x 270 x 76mm 3 57 lbs 1 7kg Quality Safety certification ISO9001 CE and CMC Working Conditions Environment temperature 41 to 95 o F 5 to 35 Environmental humidity 85 IP rating IP57 Dustproof and waterproof www burntec com www burntec com ...

Страница 18: ...onsible for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenance or repair or unauthorized modification Extech specifically disclaims any implied warranties or merchantability or fitness for a specific purpo...

Отзывы: