EA25 v3.1   8/13 

4

Operation 

Turn Power ON or OFF 

1.  Press the        button to turn the meter on or off. The meter will perform a short self-test when 

turned on. 

2.  The meter automatically turns off after 30 minutes of inactivity unless the MX/MN or Datalogging 

modes are active. 

Taking Measurements 

1.  Hold the probe in the air in the area to be tested. Do 

NOT

 immerse in liquid. 

2.  Allow adequate time for readings to stabilize. 
3.  Read the measurements on the LCD. The bargraph tracks the large digit display. 
4. The 

%RH/TEMP

 button allows the user to choose whether the temperature or the humidity is 

displayed on the large digit display. 

Selecting temperature units of measure (C/F) 

Press the 

o

C/

o

button momentarily to toggle the temperature units. 

Maximum and Minimum recording (MX/MN) 

NOTE:

 The Automatic Power OFF feature will be disabled while in the MX/MN mode. 

The MX/MN mode allows the user to view only the highest and lowest readings while taking 
measurements. (MX stands for maximum and MN for minimum).  
1. Press the 

MX/MN

 button momentarily to start the elapsed timer and begin recording. The display 

will now show only the highest reading (the MX icon will appear). 

2. Press the 

MX/MN

 button again to view the lowest reading (MN will appear). 

3. The 

SET

 button allows the user to view the elapsed time for a particular MX or MN reading in 

Minutes:Seconds

 

(59mins:59secs max)

 

or Days:Hours (99days:24hours max). 

4. To check the highest and lowest Dew Point or Wet Bulb temperatures, use the 

DEW

 and the 

WET 

buttons. 

5. The following data displays are available in the MX/MN mode: 

  Maximum and Minimum Humidity with the associated Temperature 

  Maximum and Minimum Temperature with the associated Humidity  

  Maximum and Minimum Dew Point temperature with the associated Temperature or Humidity 

  Maximum and Minimum Wet Bulb temperature with associated standard temperature or 

humidity 

6. To exit MX/MN mode, press and hold the 

MX/MN

 button (the MX/MN icons disappear). 

Data Hold 

Press the 

H

 button momentarily to freeze the displayed reading. The 'H' icon will appear on the 

upper left-hand side of the display. Press the 

H

 button again to return to normal operation (the 'H' 

hold icon will disappear). 

DEW POINT and WET BULB Temperature Display 

Dew Point and Wet Bulb temperatures are calculated by the meter based on humidity/temperature 
measurements. To view the Dew Point or Wet Bulb temperatures, press the 

DEW

 or 

WET

 button 

momentarily. The ‘DEW’ or ‘WET’ display icon will appear on the right side of the LCD and the 
temperature will display on the bottom (larger) digits. The upper (smaller) digits will display the 
regular temperature in 

o

C or 

o

F or the humidity in RH% (use the 

%RH/TEMP

 button to toggle 

between the two). 

Содержание EA25

Страница 1: ...User s Guide EasyView Digital Hygro Thermometer Datalogger Model EA25 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com ...

Страница 2: ...ce sensor Temperature thermistor Response Time Humidity 45 to 95 1 min 95 to 45 3 min Temperature Approx 2 seconds for a 10 o C change Reading Rate 2 readings per second RS 232 PC Interface Optical connection with supplied interface cable Software compatible with Windows TM XP Vista and Windows 7 systems Datalogger storage Stores 99 sets of data in manual mode 15 000 in auto mode Sampling Interval...

Страница 3: ...button 6 Humidity Temperature select button 7 SET button 8 Units C F select button 9 MEM memory button 10 ENTER button 11 READ button 12 DEW POINT temperature select left arrow button 13 UP DOWN programming keys 14 WET BULB temperature select right arrow button 15 RS 232 optical jack 16 The battery compartment is located on the rear of the instrument 9 3 6 4 7 8 5 2 1 ...

Страница 4: ... button allows the user to view the elapsed time for a particular MX or MN reading in Minutes Seconds 59mins 59secs max or Days Hours 99days 24hours max 4 To check the highest and lowest Dew Point or Wet Bulb temperatures use the DEW and the WET buttons 5 The following data displays are available in the MX MN mode Maximum and Minimum Humidity with the associated Temperature Maximum and Minimum Tem...

Страница 5: ... has been downloaded otherwise it will be lost Up to 15 000 sets of data can be recorded in automatic datalogging mode ALL of the following settings must be configured before datalogging Real Time Clock Set to the current time in hours minutes and seconds Start Time Set the time that datalogging will automatically begin Stop Time Set the time that datalogging will automatically stop Interval Time ...

Страница 6: ...ired minute that datalogging will end 5 Press to move to the two flashing digits representing seconds 6 Press or to set the desired second that datalogging will end 7 Press enter key to store the datalogging stop time and move to the next screen Datalogging Sampling Interval Setting 1 INTV interval appears along with the current interval selection in seconds 2 Use the or buttons to change the numb...

Страница 7: ...cations chart earlier in this manual Battery Replacement When the battery power falls low the BT symbol will appear on the LCD Replace the six 1 5 AAA batteries by removing the rear battery compartment screw and accessing the battery compartment Check that the polarity orientation is correct when placing the batteries in the compartment Ensure that the compartment cover is securely fastened when f...

Отзывы: