Programming Parameters - Conductivity
The Table below shows the available programming menu items P1 ~ P9. Each parameter is
explained in detail in the subsequent sections.
Parameters
Code
Selections
P1
Standard solution series selection
s ol
USA (Europe & U.S.A)
CH (China)
P2
Electrode Constant selection
0.1, 1, or 10
P3
Reference Temperature selection
77, 68, and 64
o
F (25,
20,and 18
℃
)
P4
Temperature compensation
coefficient setting
0.00 to 9.99%
P5
Electrode Constant Calibration
P6
Temperature units
℃
℉
P7
Back light display time setting
0-1-3-6min
P8
Auto power off setting
0-10-20min
P9
Restore to default factory settings
OFF / ON
Parameter P1 (Conductivity Solution Setting)
1. From the Conductivity mode, press the MODE key, the ‘P1’ icon appears on the LCD.
2. Use the CAL or the M+/RM keys to toggle through the two (2) selections: USA (for use in the
USA or Europe) and CH (for use in China).
3. Momentarily press MODE to move to the next parameter (P2), or press ENTER to return to the
normal measurement mode.
Parameter P2 (Electrode Constant ‘K’ Selection)
1. From the P2 menu, use the CAL or the M+/RM keys to select the desired electrode constant
(0.1, 1.0, or 10). The default setting is K=1.0.
2. Momentarily press MODE to move to the next parameter (P3), or press ENTER to return to the
normal measurement mode.