Specifications 

Display 

Multi-function LCD 

Maximum data points  

262,144 (Single channel or 131,072 (dual channel) 

Channels 

dual channel, TRMS AC Voltage or Current 

Sample rate 

1s to 24h selectable 

Data output 

USB port 

Current Clamp Jaw opening 

0.5” (12.7mm) 

Open input indication

 

"LO" appears on the LCD

  

(also if measured value is less than 10V/10A)

 

Low battery indication 

Empty battery symbol appears on the LCD 

Power supply 

4 AAA Batteries and 1 CR 3V memory button battery 

(CR2032 or equivalent) 

Battery life 

5 days (approximately) 

AC Adaptor 

9V, 0.5A 

Standards 

, CAT III 600V, for inside use, maximum height 

2000 meters 

Operating Temperature 

32 to 122°F (0 to 50°C) 

Operating Humidity 

<70% RH 

Storage temperature:  

-4°F to 140°F  (-20°C to 60°C) 

Storage Humidity 

< 80% relative humidity. 

Dimensions 

4.5 x 2.5 x 1.3” (114 x 63 x 34mm) 

Weight 

8.7oz. (248g) 

 

 

 

Function 

Range 

Accuracy 

Voltage (40Hz to 1kHz) 

10VAC to 600.0VAC 

±2.0%±1V 

Current (50/60Hz) 

10A to 200.0A 

±2.0%±1A 

 

 

 

 

Содержание DL160

Страница 1: ...USER GUIDE Dual Input True RMS AC Voltage Current Datalogger Model DL160 Additional User Manual translations available at www extech com ...

Страница 2: ... short circuits Do not attempt to use the instrument when your hand is wet Do not input over range during measurements Never open the battery cover during measurements Stop using the instrument when there is any structural defect or exposed metal parts Do not install substitute parts or make modifications on the meter Never replace the battery in moist areas Ensure the meter is disconnected and sw...

Страница 3: ...e is only one section filled in the battery is exhausted and should be replaced with a new one 2 AC adaptor Jack 3 ON OFF Button 4 MAX MIN Press the button one time to display the MAX maximum reading Press again to display the MIN minimum reading A third press will exit MAX MIN mode The meter will automatically exit MAX MIN mode in 10 seconds if no button is pressed 5 CH Input Jacks Insert the cur...

Страница 4: ...fore recording again 7 Remove the probes from the test circuit connect the datalogger to a PC and use the supplied software to download the data Voltage Test Leads Single Phase Connect one lead to one phase and the other lead to the Neutral line Multi phase Connect one lead to one phase and the other lead to another phase Current clamps Connect the clamp over one lead only and the clamp jaw must f...

Страница 5: ... indication Empty battery symbol appears on the LCD Power supply 4 AAA Batteries and 1 CR 3V memory button battery CR2032 or equivalent Battery life 5 days approximately AC Adaptor 9V 0 5A Standards CAT III 600V for inside use maximum height 2000 meters Operating Temperature 32 to 122 F 0 to 50 C Operating Humidity 70 RH Storage temperature 4 F to 140 F 20 C to 60 C Storage Humidity 80 relative hu...

Страница 6: ...for information on calibration and repair availability refer to the contact information below Annual calibrations should be performed to verify meter performance and accuracy Product specifications are subject to change without notice Please visit our website for the most up to date product information www extech com Contact Customer Support Customer Support Telephone List https support flir com c...

Отзывы: