TDS (Total Dissolved Solids) Calibration Procedure 

TDS CALIBRATION OPTION 1 
1. 

Allow the probe to soak in de-ionized or distilled water for 30 minutes. 

2. 

Select an appropriate TDS calibration standard. The factory default TDS conversion factor 
setting is 0.50. To change this value to better match the TDS factor of a particular calibration 
solution, refer to the Setup section. Also, refer to Appendix B (Conductivity-to-TDS Conversion 
Factors) and Appendix C (Calculating TDS Conversion Factors) for more information.  

3. 

Pour the calibration solution into two separate, clean containers to a height of 3cm. 

4. 

Power the meter and select the TDS mode using the MODE button if necessary. 

5. 

Rinse the probe in one of the calibration solution containers; gently stirring the probe. 

6. 

Dip the probe into the second calibration solution container. Tap the probe on the bottom of the 
container to remove air bubbles. Allow the probe to stabilize to the solution temperature (15 
minutes is typical). 

7. 

Press and hold the HLD/CAL button for at least 2 seconds. The TDS value and the ‘

CAL

’ icon 

on the display will flash. 

8. 

Press the MODE or HLD/CAL button to adjust the displayed TDS value so that it matches the 
standard solution value (normalized to the measured temperature). The TDS reading can only 
be adjusted to ±30% of the detected value. If detected value (displayed value) differs from the 
calibration standard by more than ±30%, the probe may need cleaning or the meter may 
require replacing. 
For example: The calibration standard is 10ppm and the detected value is 19ppm. The 
adjustable range is ±5.7ppm (19*30%). In this example the values differ over the 30% limit. 

9. 

When the CAL icon stops flashing, press the SET button momentarily to confirm the value. The 
meter will then return to the TDS measurement mode.   

 

TDS CALIBRATION OPTION 2 
TDS values are related to Conductivity; therefore the meter can be calibrated using Conductivity 
standards (as described in the Conductivity Calibration section above) and then the meter can be 
programmed with a given conversion factor. 
1. 

Perform the Conductivity Calibration as described previously. 

2. 

Select the Conductivity-to-TDS conversion factor in Setup mode (refer to the Setup section in 
this User Guide. Also, refer to Appendix B (Conductivity-to-TDS Conversion Factors) and 
Appendix C (Calculating TDS Conversion Factors) for more information.) 

3. 

Refer to the Setup section for instructions on programming the conversion factor. 

  

Maintenance 

 

Keep the meter’s measurement electrode clean

. Between measurements, rinse the electrode 

with de-ionized water. If the electrode has been exposed to a solvent immiscible in water, clean it 
with a solvent miscible in water, e.g. Ethanol, and then rinse carefully with water. 

 

Store the electrode carefully.

 Before storing, rinse carefully in de-ionized water and store dry. 

 

 

 

Office: Jl. Radin Inten II No. 62 Duren Sawit, Jakarta 13440 - Indonesia
Workshop: Jl. Pahlawan Revolusi No. 22B, Jakarta 13430 - Indonesia
Phone: 021-8690 6777 (Hunting)

Fax: 021-8690 6777

Mobile: +62 816 1740 8925

Содержание CT-3060

Страница 1: ... Meter Pen Style Water Quality Meter Office Jl Radin Inten II No 62 Duren Sawit Jakarta 13440 Indonesia Workshop Jl Pahlawan Revolusi No 22B Jakarta 13430 Indonesia Phone 021 8690 6777 Hunting Fax 021 8690 6777 Mobile 62 816 1740 8925 ...

Страница 2: ... by four 4 LR44 batteries Supplied Materials EC150 meter Four 4 LR44 button batteries Operation manual hard copy mini disk and on line availability at www extech com Battery Installation The meter is shipped with the four 4 LR44 button batteries removed The user must install the batteries before the meter can be used Refer to accompanying diagram 1 Unscrew the battery compartment cover top of mete...

Страница 3: ...million units 5 Temperature reading 6 Temperature units of measure 7 Battery strength indicator 8 Calibration icon Keypad Description 1 Power ON OFF and SET button 2 Down Arrow Data Hold and Calibration button 3 Up arrow and MODE Button Office Jl Radin Inten II No 62 Duren Sawit Jakarta 13440 Indonesia Workshop Jl Pahlawan Revolusi No 22B Jakarta 13430 Indonesia Phone 021 8690 6777 Hunting Fax 021...

Страница 4: ...lect another range as described in the Manual Range discussion in previous paragraph Set the temperature coefficient The factory default setting is 2 1 per o C this nominal value is correct for most applications Refer to the Setup section of this User Guide for details on changing this setting Also refer to Appendix D Temperature effects for more information Set the normalization reference tempera...

Страница 5: ...tive state each time it is powered down Setup Mode Parameter P1 Temperature Units Ambient Temperature and Temperature Coefficient settings 1 From the normal operating mode press and hold the SET button for at least 2 seconds until the P1 0 icon appears on the meter display 2 Press the SET button momentarily the C or F icon should now be flashing and the t ut icon abbreviation for temperature units...

Страница 6: ...uttons to select y for YES RESET or n for NO RESET 5 Press the SET button momentarily to confirm the setting 6 Press and hold the SET button for at least 2 seconds to return to the normal operation mode or press the MODE button momentarily to move to Parameter P4 see below Parameter P4 Calibration Review for Range 1 and Range 2 Concentrations 1 If continuing from Parameter P3 skip directly to step...

Страница 7: ...ir bubbles Allow the probe to stabilize to the solution temperature 15 minutes is typical 7 Press and hold the HLD CAL button for at least 2 seconds The conductivity value and the CAL icon on the display will flash 8 Press the MODE or HLD CAL buttons to adjust the displayed conductivity value so that it matches the standard solution value normalized for the measured temperature The conductivity re...

Страница 8: ...the probe may need cleaning or the meter may require replacing For example The calibration standard is 10ppm and the detected value is 19ppm The adjustable range is 5 7ppm 19 30 In this example the values differ over the 30 limit 9 When the CAL icon stops flashing press the SET button momentarily to confirm the value The meter will then return to the TDS measurement mode TDS CALIBRATION OPTION 2 T...

Страница 9: ...air meter E04 Temperature error TDS ERRORS Measurement outside of range In Manual Range mode press and hold the up arrow for 2 seconds to change range or use the Auto Range mode E04 Temperature error TEMPERATURE ERRORS E01 Temperature circuit damaged Repair meter E02 Temperature value is below allowable range or Temperature circuit damage Check again at room temperature If error persists repair me...

Страница 10: ...ating TDS Conversion Factors This meter can be calibrated using TDS calibration standard solutions The calibration requires the TDS value at a standard temperature such as 25 C Note that the TDS calibration can be performed using the conductivity calibration detailed earlier in this user guide and thereafter using a conductivity to TDS conversion factor To determine the Conductivity to TDS convers...

Страница 11: ...ctivity at T Tref Reference temperature T Sample temperature Temperature coefficient Note The correction is accurate only within a limited temperature range near T1 and T2 the greater the difference between T and Tref the higher the risk of error Calculating Temperature Coefficient Measuring the conductivity of a sample at temperature T1 close to Tref and another temperature T2 the temperature coe...

Страница 12: ...old Freezes displayed reading Automatic Temperature Compensation ATC 0 to 50 C Waterproof IP65 rated TDS Factor Selectable from 0 40 to 1 00 Temperature Coefficient Selectable from 0 to 4 0 C Normalization Temperature Reference Temperature Selectable 20 C or 25 C Basic status indicators Out of range and low battery Power Supply Four 4 LR44 button batteries Dimensions 165 x 35 x 32mm 6 5 x 1 4 x 1 ...

Отзывы: