Parameter P2: TDS Conversion Factor
1. If continuing from Parameter P1 skip directly to step 2 below. If starting from the normal
operating mode, press and hold the SET button for at least 2 seconds until the ‘
Px
’ icon appears
on the meter display (x = setup parameter number).
2. Use the MODE button to scroll to the P2 icon. The ‘
tdS
’ display will be visible above the P2.0
icon.
3. Press the SET button momentarily and the current TDS factor setting should start flashing (0.50
is the factory default setting).
4. Use the arrow buttons to change the factor; the available range is 0.40 to 1.00.
5. Press the SET button momentarily to confirm the change.
6. Press and hold the SET button for at least 2 seconds to return to the normal operation mode or
press the MODE button momentarily to move to Parameter P3 (see below).
Parameter P3: Meter Reset
This parameter can be used to restore all settings to their factory default state.
1. If continuing from Parameter P2 skip directly to step 2 below. If starting from the normal
operating mode, press and hold the SET button for at least 2 seconds until the ‘
Px
’ icon appears
on the meter display (x = setup parameter number).
2. Use the arrow buttons to scroll to the P3 icon if necessary. The ‘
rSt
’ display icon will be visible
above the P3 icon.
3. Press the SET button momentarily; a ‘
y
’ or an ‘
n
’ will be flashing.
4. Use the arrow buttons to select ‘
y
’ for YES RESET or ‘
n
’ for NO RESET.
5. Press the SET button momentarily to confirm the setting.
6. Press and hold the SET button for at least 2 seconds to return to the normal operation mode or
press the MODE button momentarily to move to Parameter P4 (see below).
Parameter P4: Calibration Review for Range 1 and Range 2 Concentrations
1. If continuing from Parameter P3 skip directly to step 2 below. If starting from the normal
operating mode, press and hold the SET button for at least 2 seconds until the ‘
Px
’ icon appears
on the meter display (x = setup parameter number).
2. Use the MODE button to scroll to the P4.0 icon if necessary. The ‘
CAL
’ display icon will be
visible above the P4.0 icon.
3. Press the SET button momentarily to view the current Range 1 Calibration Concentration. The
P4.0 icon will change to P4.1. If dashes (- - -) appear on the display this indicates that the meter
has not been calibrated thus far.
4. Press the up arrow button to move to the P4.2 display. The displayed value now represents the
Range 2 Calibration Concentration. Again, if dashed lines appear, then the meter has not been
calibrated up to this point.
5. Press and hold the SET button for at least 2 seconds to return to P4.0. Press the MODE button
momentarily to move back to parameter P1.
6. Press and SET for at least 2 seconds to return to normal operation mode.
Office: Jl. Radin Inten II No. 62 Duren Sawit, Jakarta 13440 - Indonesia
Workshop: Jl. Pahlawan Revolusi No. 22B, Jakarta 13430 - Indonesia
Phone: 021-8690 6777 (Hunting)
Fax: 021-8690 6777
Mobile: +62 816 1740 8925