Extech Instruments AN310 Скачать руководство пользователя страница 4

AN310-en-GB_V1.5   11/15 

4

SET 

HOLD 

MAX/MIN 

 

U/mode 

AVG 

 

L/mode 

 

U/mode 

AVG 

 

Brief

 

Keypad

 

Description

 

 

Press

 

momentarily

 

to

 

switch

 

the

 

meter

 

ON

 

or

 

OFF.

 

Press

 

and

 

hold

 

this

 

button

 

for

 

2

 

seconds

 

to

 

enter

 

or

 

exit

 

the

 

Setup

 

mode

 

(refer

 

to

 

Setup

 

mode

 

section

 

for

 

details).

 

 

o

 

Press

 

momentarily

 

to

 

freeze

 

the

 

displayed

 

reading

 

(HOLD

 

icon

 

switches

 

ON);

 

press

 

again

 

momentarily

 

to

 

exit

 

the

 

HOLD

 

mode

 

(HOLD

 

icon

 

switches

 

OFF)

 

o

 

Press

 

and

 

hold

 

for

 

2

 

seconds

 

to

 

display

 

the

 

MIN

 

(minimum)

 

reading;

 

press

 

and

 

hold

 

again

 

to

 

display

 

the

 

MAX

 

(maximum)

 

reading;

 

Press

 

and

 

hold

 

a

 

third

 

time

 

to

 

exit

 

o

 

This

 

button

 

is

 

also

 

used

 

in

 

the

 

AVERAGE

 

mode

 

(for

 

recording

 

readings

 

one

 

at

 

a

 

time

 

or

 

for

 

a

 

‘timed’

 

recording

 

session)

 

and

 

in

 

the

 

SETUP

 

mode

 

(for

 

configuring

 

the

 

meter).

 

Please

 

review

 

the

 

AVERAGE

 

and

 

SETUP

 

mode

 

sections

 

for

 

details

 

 

 

Press

 

momentarily

 

to

 

toggle

 

RH

 

and

 

Air

 

Velocity

 

readings.

 

Press

 

and

 

hold

 

for

 

2

 

seconds

 

to

 

access

 

the

 

Multi

Point

 

Average

 

and

 

Timed

 

Average

 

modes

 

(the

 

AVG

 

icon

 

switches

 

ON).

 

Detailed

 

instructions

 

for

 

these

 

modes

 

are

 

provided

 

later

 

in

 

this

 

guide.

 

In

 

the

 

Setup

 

mode

 

press

 

to

 

select

 

a

 

category

 

or

 

to

 

increase

 

a

 

value

 

(refer

 

to

 

Setup

 

mode

 

later

 

in

 

this

 

guide)

 

 

 

Press

 

to

 

toggle

 

the

 

lower

 

display

 

between

 

air

 

temperature,

 

air

 

volume,

 

wet

 

bulb

 

temperature,

 

and

 

dew

 

point

 

temperature

 

readings;

 

Also

 

used

 

in

 

the

 

Multi

Point

 

Average,

 

Timed

 

Average,

 

and

 

Setup

 

modes;

 

refer

 

to

 

the

 

AVERAGE

 

and

 

SETUP

 

modes

 

sections

 

later

 

in

 

this

 

guide

 

 

 

   

 

      

 

 

+

 

With

 

the

 

meter

 

switched

 

OFF,

 

press

 

and

 

hold

 

these

 

two

 

buttons

 

to

 

disable

 

Auto

 

Power

 

OFF.

 

Detailed

 

AUTO

 

POWER

 

OFF

 

(Sleep

 

mode)

 

instructions

 

are

 

provided

 

later

 

in

 

this

 

guide

 

 

contact:

[email protected]

www.ponpe.com

Содержание AN310

Страница 1: ...USER GUIDE Large Vane CFM CMM Anemometer Psychrometer Model AN310 contact sale ponpe com www ponpe com ...

Страница 2: ...se damage to the meter and personal injury Read and understand this user manual before operating the meter Inspect the condition of the probe and the meter for any damage before operating the meter Repair or replace damage before use If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired This device should not be made availab...

Страница 3: ...arrow button 4 L mode and down arrow button 5 ON OFF Power and SET button 6 HOLD MAX MIN and left arrow Enter button 7 Unit of measure symbol for lower display row 8 Unit of measure symbol for upper display row 9 Air Velocity Vane Meter Description rear 10 Air temperature and humidity sensors 11 Battery compartment contact sale ponpe com www ponpe com ...

Страница 4: ...for configuring the meter Please review the AVERAGE and SETUP mode sections for details Press momentarily to toggle RH and Air Velocity readings Press and hold for 2 seconds to access the Multi Point Average and Timed Average modes the AVG icon switches ON Detailed instructions for these modes are provided later in this guide In the Setup mode press to select a category or to increase a value refe...

Страница 5: ...3 Optional Cone or Funnel attachment icon 4 Units of measure for duct area programming 5 Vane indicator flashes very briefly on power up 6 Display HOLD icon 7 MAX mode icon 8 MIN mode icon 9 Recording icon 10 Elapsed Timer icon 11 AVG mode icon 12 Battery status icon 13 Unit of measure icons Full list shown Available units of measure icons vary by model contact sale ponpe com www ponpe com ...

Страница 6: ... CFM cubic feet per minute or CMM cubic meters per minute Use the L mode button for lower display and the U mode button for upper display to scroll through the displayed reading options 6 To change between Imperial and Metric units of measure refer to the SETUP Mode section 7 To measure the Air Volume of a duct first measure the duct s area see the Appendix for area calculations and then enter the...

Страница 7: ... will appear at the top of the LCD display indicating that the readings shown are the lowest measured 2 Press and hold the MAX MIN button again for 2 seconds to display the maximum readings encountered The MAX icon will appear on the LCD 3 Press and hold this button again for 2 seconds to return to the normal mode of operation the MIN and MAX icons should switch OFF 4 While viewing the MAX or MIN ...

Страница 8: ...ter the P1 0 UNIT menu 2 Use either arrow button to switch between metric and imperial units The available units are air velocity m s fpm temperature C F air volume CMM CFM and area size cm inch 3 Press HOLD momentarily to return to the P1 0 display 4 Press either arrow button to move to the P2 0 AREA option and follow the steps below P2 0 Area Size Selection enter square inches or square centimet...

Страница 9: ...ver dispose of used batteries or rechargeable batteries in household waste As consumers users are legally required to take used batteries to appropriate collection sites the retail store where the batteries were purchased or wherever batteries are sold Disposal Do not dispose of this instrument in household waste The user is obligated to take end of life devices to a designated collection point fo...

Страница 10: ...Circuit Custom LSI microprocessor Display Dual function 13 mm 0 5 4 digit LCD Sampling rate 1 reading per second approx Air velocity flow sensor Conventional angled vane arms with low friction ball bearing Sensors NTC type precision thermistors Automatic Power off Auto shut off after 20 minutes to preserve battery life sleep mode Operating Temperature 0 C to 50 C 32 F to 122 F Storage Temperature ...

Страница 11: ... error Check temperature or RH error solutions Air velocity E03 Velocity exceeds the high range limit Test the meter using an air velocity that is known to be within the specified range limits Send the unit to Extech for repair if the E03 error message persists Air volume E03 Reading exceeds the meter s display limit Check that the AREA setting is correct in the Setup Mode E04 Air velocity error R...

Страница 12: ...196 87 1 944 3 6 2 24 1 ft min 0 00508 1 0 00987 0 01829 0 01138 1 knot 0 5144 101 27 1 1 8519 1 1523 1 km h 0 2778 54 69 0 54 1 0 6222 1 MPH 0 4464 87 89 0 8679 1 6071 1 Copyright 2014 2015 FLIR Systems Inc All rights reserved including the right of reproduction in whole or in part in any form www extech com Width W Height H Area A Width W x Height H CFM ft 3 min Air Velocity ft min x Area ft 2 C...