background image

EA33-en-GB_v2.5 7/18

4

MODE 2   Elapsed Timer with Hold 

In this mode, measurements are made while the countdown timer is running. When the timer 
reaches zero, the last measurement is held on the display (see diagram). 
1. Press the SET key;  SEt 01  will display. 
2. Press the down arrow key to change to the  SEt 02  parameter. 

-

 

3. Press the ENTER key to call up the timer. 
4. Use the up/down arrow keys to set the desired seconds.  
5. Press the ENTER key to program the number of minutes. 
6. Use the up/down arrow keys to set the minutes. 
7. Press the ENTER key to start the countdown timer. When the timer 

counts down, the last reading will be held on the display. 

8. Press the ENTER key to exit this mode. 

MODE 3 - Real Time Clock Setting 

1.

 

2. Press the d

 

right of the LCD. 

3.

 

4.

 

5. Repeat the previous two steps for the  minutes ,  hours , and  day  settings. 
6. Press the ENTER key when finished. 
7. Use the TIME key to toggle between minutes-seconds (m-s) and day-hour (D-H) views. 

MODE 4 - Integral Illuminance Mode 

Integral Illuminance is a luminous energy measurement based on the rate of flow of flux (expressed 
in lumen-seconds). The Model EA33 can also measure any photometric quantity on a time 
dependent basis. For example, the illuminance could be integrated over time to yield lux-hours (lx-h) 
or foot candle-hours (fc-h). 
1. Press th

SEt 0

 

2.

 

 

3. Press the ENTER key to start the Integral Illuminance 

measurement. The meter will set the first measurement as a 
reference. 

-

-

appear and the timer 

will start (the cycle number appears on the lower left of the LCD). 

4. If measurements exceed the reference value, the display will 

 Note that the timer will also 

halt. 

5. The maximum indication of integral illuminance is 999999. When 

this maximum is exceeded, the integration number (Cycle no. at 
lower left of LCD) can be checked. The maximum number of 
integration cycles is 99. 

6. Press the ENTER key to exit this mode of operation. 

www.

.com

[email protected]

1.800.561.8187

Содержание 793950411339

Страница 1: ...User Manual EasyView Light Meter with Memory Model EA33 www com information itm com 1 800 561 8187 ...

Страница 2: ...Intensity cd Comparator button 8 READ Right arrow button 9 Average reading tally display 10 Power ON OFF button 11 Set button 12 Average button 13 Relative Percent button 14 Light sensor 15 Left arrow Memory button 16 Up Down arrow buttons Display 1 Relative Mode 2 Comparator Mode 3 Average 4 Data Hold 5 Relative 6 Candela 7 Relative 8 fc lx hours Integral Illuminance 9 Stray light measurement 10 ...

Страница 3: ...e Description SEt 01 Stray Light Cancels out stray light from light under test SEt 02 Time Hold Meter measures while timer counts down Reading is held on LCD when counter 0 SEt 03 rtC Set the day and time for internal clock SEt 04 Int Lumen seconds lux hour and fc hour measurements SEt 05 COMP Measurements are compared to high and low limits MODE 1 Stray Light Inhibit This feature allows the user ...

Страница 4: ...hed 7 Use the TIME key to toggle between minutes seconds m s and day hour D H views MODE 4 Integral Illuminance Mode Integral Illuminance is a luminous energy measurement based on the rate of flow of flux expressed in lumen seconds The Model EA33 can also measure any photometric quantity on a time dependent basis For example the illuminance could be integrated over time to yield lux hours lx h or ...

Страница 5: ...ement exceeds the high limit the plus sign will appear If the low limit is exceeded the minus sign will appear An audible beep will sound when a limit has been reached 14 Press the ENTER key to exit the Comparator mode 99 Point Averaging Illuminance is the amount of visible light incident upon a specific surface area Since the EA33 sensor area is small the Average Mode can be used to measure the i...

Страница 6: ...er turned on a data record is recorded The and an audible tone will sound 2 The counter on the lower left of the LCD will show the number of stored readings 3 The total memory size is 50 records READING STORED RECORDS 1 Press the READ key to enter the number for the displayed data point 2 Use the up down arrow keys to scroll through the stored readings 3 Press the ENTER key to exit this mode of op...

Страница 7: ...is feature 1 Turn the meter OFF 2 Press and hold the HOLD key while turning the meter ON 3 The AUTO POWER OFF feature will now be disabled 4 Note that AUTO POWER OFF is re enabled each time the meter is turned on 5 Also note that AUTO POWER OFF is always disabled in the Integral Illuminance mode www com information itm com 1 800 561 8187 ...

Страница 8: ...dispose of used batteries or rechargeable batteries in household waste As consumers users are legally required to take used batteries to appropriate collection sites the retail store where the batteries were purchased or wherever batteries are sold Disposal Do not dispose of this instrument in household waste The user is obligated to take end of life devices to a designated collection point for th...

Страница 9: ...hoto detector Silicon photo diode with spectral response filter Operating conditions Temperature 0 to 40o C 32 to 104o F Humidity 80 RH Storage conditions Temperature 10 to 50o C 14 to 140o F Humidity 80 RH Meter Dimensions 150 x 72 x 35mm 5 91 x 2 8 x 1 4 Detector Dimensions 92 x 60 x 29mm Weight Approx 11 3 oz 320g with battery Sensor lead length 150cm Low battery indication Battery symbol appea...

Страница 10: ...33 en GB_v2 5 7 18 10 Appendix Typical Light Levels Spectral Sensitivity 0 0 2 0 4 0 6 0 8 1 1 2 Wavelength in Nonometers 0 0 2 0 4 0 6 0 8 1 1 2 Light Meter CIE www com information itm com 1 800 561 8187 ...

Отзывы: