V01.02.01 7/04
7
Diode Test
Test current of 0.3mA maximum,
open circuit voltage 1.5V DC typical
Continuity Check
Audible signal will sound if the
resistance is less than 150
Ω
(approx.), test current <0.7mA
Temperature Sensor
Requires type K thermocouple
Input Impedance
>7.5M
Ω
(VDC & VAC)
AC Response
True rms
ACV Bandwidth
50Hz to 1kHz
Display
4000 count backlit liquid crystal
Overrange indication
“OL” is displayed
Auto Power Off
15 minutes (approximately)
Polarity
Automatic (no indication for
positive); Minus (-) sign for negative
Measurement Rate
2 times per second, nominal
Low Battery Indication
“ ”
is displayed if battery voltage
drops below operating voltage
Battery
One 9 volt (NEDA 1604) battery
Fuses
mA, µA ranges; 0.5A/250V fast blow
A range; 20A/250V ceramic fast
blow
Operating Temperature
41ºF to 104ºF (5ºC to 40ºC)
Storage Temperature
-4
o
F to 140
o
F (-20
o
C to 60
o
C)
Operating Humidity
Max 80% up to 87ºF (31ºC)
decreasing linearly to 50% at
104ºF(40ºC)
Storage Humidity
<80%
Operating Altitude
7000ft. (2000meters) maximum.
Weight
0.753lb (342g) (includes holster).
Size
7.36” x 3.2” x 2.0” (187 x 81 x
50mm) (includes holster)
Safety
For indoor use and in accordance with
the requirements for double insulation to
IEC1010-1 (1995): EN61010-1 (1995)
Overvoltage Category III 600V and
Category II 1000V, Pollution Degree 2.