42535   Version 4.0   01/05 

Type K Thermocouple Temperature Measurements 

1. 

Connect a K-Type temperature probe into the socket on the front of the handle. 

2. 

Press the Mode button until 

PRB

 appears on the display. 

3. 

Read the temperature on the display. 

4. 

To see the minimum temperature recorded using the probe, press and hold the 

ºC/ºF

 

button. 

5. 

To see the maximum temperature recorded, press and hold the 

LOCK

 button. 

6. 

In this mode, the meter will automatically power off if left idle for more than 12 minutes. 

Type K Measurement Considerations 

When a Type K Thermocouple is first connected to the meter and the thermocouple and 
meter temperatures are different, several minutes should be allotted so that the temperature 
of the socket matches the temperature of the thermocouple.  

High and Low Alarms 

1. 

To set the high alarm press the Mode button until 

HAL

 

(high alarm) appears in the display. 

Use the Up and Down arrow buttons to set the alarm temperature. 

2. 

To set the Low alarm, press the Mode button until 

LAL

 (low alarm) appears in the 

display. Use the Up and Down buttons to set the alarm temperature. 

3. 

When the measured temperature exceeds the settings, the alarm will sound and the “

HI

” 

or “

LOW

” icon will be displayed              

Emissivity adjustment 

The emissivity is factory set to 0.95, which covers 90% of all applications. However, the 
closer the meter's emissivity setting is to the actual emissivity of the object under test, the 
more accurate the temperature measurements will be. To calculate the emissivity of an 
object: Measure an object's temperature using a Type K Thermocouple and note the reading. 
Measure the object again using the IR sensor. Now, press the Mode button until the E symbol 
(emissivity) appears on the screen. Using a small screwdriver or paper clip, press and hold 
down the Emissivity button. Use the UP and Down arrow buttons to set the Emissivity. 
Release the Emissivity button. Press the Mode button to confirm the setting. 

Back Light Display 

To turn the Back Light On/Off, press and hold the measuring trigger. While continuing to pull 
the trigger, press the 

LOCK

 button to turn the light on/off. When the back light is on, the 

 

icon will appear on the display. 

Error Message 

If the 

 symbol is displayed, it is necessary to reset the thermometer. To reset it, turn the 

instrument off, remove the battery and wait for a minimum of one minute, reinsert the battery 
and turn on. If the error message remains repair/replacement of the meter is necessary. 

Battery Replacement 

This symbol 

 indicates that the battery is OK. When the low battery symbol 

 

appears on 

the LCD, measurements are still possible but the meter’s 2 AAA batteries should be replaced. 

This symbol 

 indicates that the battery is exhausted and no measurements can be taken. 

The hinged battery compartment is located on the bottom of the meter's handle. Open the 
compartment by sliding the battery compartment cover down and lifting it from the bottom. 
Replace the batteries and close the battery compartment cover. 

 

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

FAX 781.665.0780 - TestEquipmentDepot.com 

Содержание 42535

Страница 1: ...mocouple functionality is also built in Proper use and care of this meter will provide years of reliable service Safety 1 Use extreme caution when the laser beam is ON 2 Do not point the beam toward anyone s eyes 3 Be careful not to let the beam strike the eye from a reflective surface 4 Do not use the laser near explosive gases or in other potentially explosive areas Test Equipment Depot 800 517 ...

Страница 2: ...tton to turn the laser pointer on or off When the laser is on the icon will appear on the top left of the display 5 While in the HOLD mode press the ºC ºF button to change the temperature units 6 Momentarily press the Mode button to step through and view the MAX maximum MIN minimum dIF difference between the max and min AVG average HAL high alarm LAL low alarm PRB type k probe or E emissivity 7 Wh...

Страница 3: ...erature measurements will be To calculate the emissivity of an object Measure an object s temperature using a Type K Thermocouple and note the reading Measure the object again using the IR sensor Now press the Mode button until the E symbol emissivity appears on the screen Using a small screwdriver or paper clip press and hold down the Emissivity button Use the UP and Down arrow buttons to set the...

Страница 4: ...ions Display 0 31 7 87mm 4 digit LCD display Sample rate 1 sec approx Laser power Laser power less than 1mW red 645 660nm Operating Temperature 32 F to 122 F 0 C to 50 C Operating Humidity Max 80 RH Power Supply 2 AAA batteries Battery Life 180 hrs typical without Laser and Back Light Weight 6 3oz 179g Size 6 9 x 1 5 x 2 8 175 x 39 x 72 mm Infrared Thermometer Specifications Range Resolution 76 to...

Отзывы: