background image

42535   Version 4.0   01/05 

Meter Description 

 
1. LCD 

Display 

2. Emissivity 

button 

3. Down 

button/°F°C 

4. Mode 

button 

5. Up 

button/Lock 

6. Battery 

cover 

7. Infrared 

lens 

8. Laser 

pointer 

9. Measuring 

trigger 

10. Thermocouple 

input 

 

Operating Instructions 

Infrared Temperature 

1. 

Point the meter toward the surface to be measured. 

2. 

Pull and hold the trigger to turn the meter on and begin testing. “

SCAN

” and a blinking 

        icon will appear in the display while the trigger is held. 

3. 

Release the trigger and “

HOLD

” will appear. The last reading will remain in the display 

for 60 seconds and then the meter will shut off. 

4. 

While measuring, press the 

ºC/ºF

 button to turn the laser pointer on or off. When the 

laser is on, the   icon will appear on the top left of the display. 

5. 

While in the HOLD mode, press the 

ºC/ºF

 button to change the temperature units. 

6. 

Momentarily press the 

Mode

 button to step through and view the 

MAX

 (maximum), 

MIN

 

(minimum), 

dIF

 (difference between the max and min), 

AVG

 (average), 

HAL

 (high 

alarm), 

LAL

 (low alarm), 

PRB

 (type k probe) or 

E

 (emissivity). 

7. 

While in the HOLD mode, momentarily press the 

LOCK

 button to allow the meter to run 

continuously for 60 minutes without holding the trigger. The “

LOCK

” and “

SCAN

” icons 

will appear in the display. Press the 

LOCK

 button again to exit Lock mode. 

 

 

 

 

 

 

 

 

 

0.5 

  1.0 

  2.0 

5.5 

11 

22 

Diameter of Spot (inches) 

Distance to Object (inches) 

Laser Pointer 

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

FAX 781.665.0780 - TestEquipmentDepot.com 

Содержание 42535

Страница 1: ...mocouple functionality is also built in Proper use and care of this meter will provide years of reliable service Safety 1 Use extreme caution when the laser beam is ON 2 Do not point the beam toward anyone s eyes 3 Be careful not to let the beam strike the eye from a reflective surface 4 Do not use the laser near explosive gases or in other potentially explosive areas Test Equipment Depot 800 517 ...

Страница 2: ...tton to turn the laser pointer on or off When the laser is on the icon will appear on the top left of the display 5 While in the HOLD mode press the ºC ºF button to change the temperature units 6 Momentarily press the Mode button to step through and view the MAX maximum MIN minimum dIF difference between the max and min AVG average HAL high alarm LAL low alarm PRB type k probe or E emissivity 7 Wh...

Страница 3: ...erature measurements will be To calculate the emissivity of an object Measure an object s temperature using a Type K Thermocouple and note the reading Measure the object again using the IR sensor Now press the Mode button until the E symbol emissivity appears on the screen Using a small screwdriver or paper clip press and hold down the Emissivity button Use the UP and Down arrow buttons to set the...

Страница 4: ...ions Display 0 31 7 87mm 4 digit LCD display Sample rate 1 sec approx Laser power Laser power less than 1mW red 645 660nm Operating Temperature 32 F to 122 F 0 C to 50 C Operating Humidity Max 80 RH Power Supply 2 AAA batteries Battery Life 180 hrs typical without Laser and Back Light Weight 6 3oz 179g Size 6 9 x 1 5 x 2 8 175 x 39 x 72 mm Infrared Thermometer Specifications Range Resolution 76 to...

Отзывы: