background image

  V1.5  1/03 

WARRANTY 

EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in 
parts and workmanship for one year from date of shipment (a six month limited warranty 
applies on sensors and cables). If it should become necessary to return the instrument for 
service during or beyond the warranty period, contact the Customer Service Department at 
(781) 890-7440 ext. 210 for authorization. 

A Return Authorization (RA) number must be 

issued before any product is returned to Extech.

 The sender is responsible for shipping 

charges, freight, insurance and proper packaging to prevent damage in transit. This warranty 
does not apply to defects resulting from action of the user such as misuse, improper wiring, 
operation outside of specification, improper maintenance or repair, or unauthorized 
modification. Extech specifically disclaims any implied warranties or merchantability or fitness 
for a specific purpose and will not be liable for any direct, indirect, incidental or consequential 
damages. Extech's total liability is limited to repair or replacement of the product. The 
warranty set forth above is inclusive and no other warranty, whether written or oral, is 
expressed or implied. 

Safety Precautions 

1.  Make sure any covers or battery doors are properly closed and secured. 
2.  Remove the battery from the meter if the meter is to be stored for long periods. 
3.  Touching the thermocouple to a device which is powered electrically can cause 

damage to the meter and may pose a shock hazard. 

4.  To avoid damage do not make temperature measurements in microwave ovens. 

Specifications 

Display 

3 ½ digit (1999 Count) LCD (Liquid Crystal Display) 

Temperature Scale 

ºC or ºF (user-selectable) 

Thermocouple type 

Model 42311: type J; Model 42312: type K 

Measurement and 

Model 42312:  -40º to 2000ºF (-40º to 1350ºC) 

Source Ranges 

Model 42311:  -40º to 1400ºF (-40º to 760ºC)  

Resolution 

1ºC/F or 0.1ºC/F (< 200ºF/C) 

Accuracy 

±(0.1% rdg + 2ºF) or ±(0.1% rdg + 1ºC) 

Temp. Coefficient 

0.1 times the applicable accuracy spec per ºC or ºF from 
32º to 64ºF and 82º to 122ºF (0º to 18ºC and 28º to 50ºC) 

Input Protection 

24VDC or 24VAC RMS maximum input voltage 

Display Update 

2.5 times per second 

Input Connector 

Accepts standard subminiature thermocouple connectors 

Battery 

9V battery (200 hour typical battery life) 

Dimensions/Weight 

7.5x3.54x2.1 inches (190x90x53 mm) / 1 lb (440g) 

Содержание 42311

Страница 1: ...Calibrator Thermometer Models 42311 Type J 42312 Type K Introduction Congratulations on your purchase of Extech model 42311 or 42312 Digital Thermometer and Calibrator Properly used this meter will provide many years of reliable service ...

Страница 2: ...nt of the product The warranty set forth above is inclusive and no other warranty whether written or oral is expressed or implied Safety Precautions 1 Make sure any covers or battery doors are properly closed and secured 2 Remove the battery from the meter if the meter is to be stored for long periods 3 Touching the thermocouple to a device which is powered electrically can cause damage to the met...

Страница 3: ...ax hold Backlit display Press the Hold key for 2 seconds to toggle the Backlight ON and OFF The backlight will stay on until you press the Hold key for 2 seconds again or turn the meter off Data Hold The hold function freezes the displayed reading for later viewing Press the HOLD key momentarily to activate the Data Hold feature Press again to exit the Hold mode Operation WARNING To avoid electric...

Страница 4: ...g to the temperature display which can be routed to a panel meter recorder datalogger controller another calibrator etc 3 Adjust the Coarse knob to dial in even closer to the desired output temperature 4 Finally adjust the Fine knob for the exact temperature output Maintenance Cleaning and Storage Periodically wipe the case with a damp cloth and mild detergent do not use abrasives or solvents If t...