Measure and Source 

MEASURE (Input)  

In this mode, the unit will measure voltage or current. 
1. 

Select Voltage or Current on the Function switch. 

2. 

Turn the meter ON 

3. 

Press the MODE button to select MEASURE 

4. 

Press the UNIT button to select mA or % if current is selected. 

5. 

Connect the Calibration Cable to the meter. 

6. 

Connect the Calibration Cable to the device or circuit under test. 

7. 

Read the measurement on the LCD display. 

SOURCE (Output)  

In this mode, the unit will source (output) voltage or current. 
1. 

Select Voltage or Current on the Function switch. 

2. 

Turn the meter ON 

3. 

Press the MODE button to select SOURCE 

4. 

Press the UNIT button to select V or mV if voltage is selected or to select mA or % if current 
is selected. 

5. 

Connect the Calibration Cable to the meter. 

6. 

Connect the Calibration Cable to the device or circuit under test. 

7. Use 

the 

 

buttons to select the desired output value.

 

Use the LCD display to verify the 

output level. Alternatively, use the stored calibration values in memory as described in the 
MEM button section. 

8. 

For the -25% to 125% output range the output is 0 to 24mA. 


Display 

-25% 

0% 

25% 50%  75% 

100% 

125% 

mA 
output 

0mA 

4mA 

8mA 12mA 16mA 

20mA 

24mA 

  

NOTE

: The “

SOURCE”

 icon will blink when the output value has not reached a stable level. 

The common cause for the “

SOURCE”

 icon to remain blinking is that the load impedance is 

too high in the current mode or too low in the voltage mode.  

 

 

 

 

Содержание 412355A Oyster Series

Страница 1: ...User s Guide Model 412355A Current Voltage Calibrator ...

Страница 2: ...C adaptor jack 9 Battery compartment 10 Neck strap holders 11 Function switch Battery Replacement When the LOW BAT message appears on the LCD replace the 9V battery as soon as possible 1 Open the calibrator s lid as far as possible 2 Open the battery compartment using a coin at the arrow indicator 3 Replace the battery and close the cover You as the end user are legally bound Battery ordinance to ...

Страница 3: ...f activated FUNCTION SLIDE SWITCH Slide the Function switch on the side of the meter to the desired function Voltage or Current MODE Press the MODE button to select either SOURCE output or MEASURE input UNIT BUTTON Press the UNIT button to select 1 Voltage mode mV or V in source 2 Current mode mA or in source or measure OUTPUT ADJUST BUTTONS Press the buttons to increase or decrease the output vol...

Страница 4: ...n the current mode or too low in the voltage mode Storing values into memory 1 Select and display a memory location M1 to M5 2 Press the buttons to adjust the display to the desired source value 3 Press and hold the MEM button for 2 seconds The displayed value will be stored into the displayed memory location Default memory values Five common source values are permanently programmed in memory as t...

Страница 5: ...ODE button to select SOURCE 4 Press the UNIT button to select V or mV if voltage is selected or to select mA or if current is selected 5 Connect the Calibration Cable to the meter 6 Connect the Calibration Cable to the device or circuit under test 7 Use the buttons to select the desired output value Use the LCD display to verify the output level Alternatively use the stored calibration values in m...

Страница 6: ...ting Altitude 2000meters 7000ft maximum Dimensions 96 x 118 x 45mm 3 8 x 4 7 x 1 8 folded Weight 340g 12 oz Accessories Supplied 9V battery AC adaptor and calibration cable with spade lugs Range Specifications Mode Function Range Resolution Accuracy of reading Measure Voltage 0 to 2000mV 1mV 0 075 1 digit or 3 digits whichever is greater 2 00 to 20 00V 0 01V Current 0 00 to 50 00mA 0 01mA 25 0 to ...

Отзывы: