Extech Instruments 407780 Скачать руководство пользователя страница 3

Model

 

407780

EN

 

v2.3

   

1/14

 

 

3

Refer to the diagram on the facing page: 

1.  1/2-inch microphone (shown with windscreen) 
2. LCD 

display 

3. Power 

Switch 

4.  Leq / SEL / SPL select switch 
5.  RUN / Pause switch 
6.  Use the UP/DOWN arrow keys to change the measurement preset time. 
7.  A / C Frequency weighting select switch 
8.  FAST / SLOW / IMPULSE time weighting select switch. 

FAST

: 125ms  

SLOW

: 1 second 

IMPULSE

: 35ms with slow decay 

9.  Press to select MAX MIN recording. Press again to step through MAX and MIN values. 

Press and hold for 3 seconds to erase the MAX and MIN values and exit this mode. 

10. RECORD / ERASE 

Record standby

: The RECORD symbol appears in standby mode. 

Recording

: The RECORD symbol flashes once per second while recording. 

Erase data

: Press & hold the RECORD/ERASE key for 3 seconds to erase data (the entire LCD 

will flash three times).  

11. Real time clock with calendar 
12. Measuring time selection for Leq and SEL 
13. RS-232 interface connector (Located on bottom, not shown) 
14. AC output terminal (located on side, not shown) 

2 Vrms at 130dB 
Output impedance 600

 Max. 

Output signal uses standard 3.5mm phono jack (supplied). Signal on tip; ground on sleeve. 

15. DC output terminal (located on side, not shown) 

Output: 10mV/dB 
Output impedance 100

 max. 

Output signal uses standard 3.5mm phono jack (supplied). Signal on tip and ground on sleeve. 

16. CAL (calibration) potentiometer (located on side, not shown) 
17. External DC 6V power supply terminal (located on side, not shown) 
18. Tripod mounting screw (located on rear, not shown) 
19. Battery cover (located on rear, not shown) 

 

Содержание 407780

Страница 1: ...User s Guide Integrating Sound Level Datalogger Model 407780...

Страница 2: ...MAX L and MIN L measurements Response time FAST SLOW and IMPULSE and frequency weighting A and C are programmable The analog output AC and DC can be used with headphones chart recorders or other devic...

Страница 3: ...bol flashes once per second while recording Erase data Press hold the RECORD ERASE key for 3 seconds to erase data the entire LCD will flash three times 11 Real time clock with calendar 12 Measuring t...

Страница 4: ...of sound pressure level 10 Data recording 11 Measuring value 12 Memory full 13 Unit 14 Frequency weighting A C 15 IMPULSE time weighting 16 When the TIME display icon switches on the real time clock...

Страница 5: ...orm a manual recording select the 24 hour interval Change the run time by pressing the INTEG TIME button The run time will appear in the middle of the LCD Press the or button to select the preset time...

Страница 6: ...e To perform a manual recording select the 24 hour interval Change the run time by pressing the INTEG TIME button The run time will appear in the middle of the LCD Press the or button to select the pr...

Страница 7: ...to capture noises that occur very quickly In FAST mode the meter responds in 125ms Select the SLOW Mode meter responds in 1s to monitor a sound source that has a reasonably consistent noise level or t...

Страница 8: ...the arrow keys to set the correct month 4 Repeat step 3 to set the day hour minute and second 5 To confirm the new date and time press the DATE TIME button Note that the 407780 has a back up battery N...

Страница 9: ...th a dry cloth Do not use abrasives or solvents Analog output The 407780 has two analog output jacks located on its right side one for AC and one for DC For DC the meter transmits 10mv dB For AC the f...

Страница 10: ...and the measurement time in the supplied software The meter can now record remotely disconnected from the PC at the rate selected 4 Press the RECORD ERASE key to put the meter in Record Standby The RE...

Страница 11: ...ay 4dB per step over 100dB range with 50 segments Bargraph update rate 16 times per second Display warnings OVER and UNDER range status indicators Analog outputs DC 10mV per dB AC 2Vrms full scale Pow...

Страница 12: ...al or consequential damages FLIR s total liability is limited to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is express...