Specifications 

General Specifications 

Applicable Standards 

IEC 651 / 804 type 2 and ANSI S1.4 type 2 

Measurement types 

SPL, SEL, Leq, MAX-L, and MIN-L 

Measurement range 

30 to 130dB 

Frequency Range 

31.5Hz to 8kHz 

Frequency weighting 

‘A’ and ‘C’ 

Response time   

FAST, SLOW, and IMPULSE selections 

Linearity range   

100dB 

Display resolution 

0.1dB 

Accuracy  

 

±1.5db (94dB @ 1kHz) 

Microphone 

 

0.5” Electret Condensor 

Numeric Display   

4-digit LCD 

Display update rate 

Twice per second 

Bargraph Display  

4dB per step (over 100dB range) with 50 segments 

Bargraph update rate 

16 times per second 

Display warnings  

‘OVER’ and ‘UNDER’ range status indicators 

Analog outputs   

DC: 10mV per dB; AC: 2Vrms full scale 

Power supply  

 

Four (4) 1.5V ‘AA’ batteries (optional AC adaptor) 

Battery life 

 

20 hours (approx.) 

Operating conditions 

41 to 104

o

F (5 to 40

o

C); 90% Relative Humidity 

Storage conditions 

14 to 140

o

F (-10 to 60

o

C); 75% Relative Humidity 

Dimensions 

 

10.4 x 2.8 x 0.8" (265 x 72 x 21mm) 

Weight 

  Approx. 

10.9 

oz. 

(310g) 

 

Supplied Material 

Adjustment screwdriver 

Batteries (four ‘AA’ 1.5V cells) 

Operation Manual 

CD with Windows

TM

 Software and software manual 

RS-232 Interface cable  

Windscreen 

3.5mm plug for analog output access 

 

Содержание 407780

Страница 1: ...User s Guide Integrating Sound Level Datalogger Model 407780...

Страница 2: ...MAX L and MIN L measurements Response time FAST SLOW and IMPULSE and frequency weighting A and C are programmable The analog output AC and DC can be used with headphones chart recorders or other devic...

Страница 3: ...bol flashes once per second while recording Erase data Press hold the RECORD ERASE key for 3 seconds to erase data the entire LCD will flash three times 11 Real time clock with calendar 12 Measuring t...

Страница 4: ...of sound pressure level 10 Data recording 11 Measuring value 12 Memory full 13 Unit 14 Frequency weighting A C 15 IMPULSE time weighting 16 When the TIME display icon switches on the real time clock...

Страница 5: ...orm a manual recording select the 24 hour interval Change the run time by pressing the INTEG TIME button The run time will appear in the middle of the LCD Press the or button to select the preset time...

Страница 6: ...e To perform a manual recording select the 24 hour interval Change the run time by pressing the INTEG TIME button The run time will appear in the middle of the LCD Press the or button to select the pr...

Страница 7: ...to capture noises that occur very quickly In FAST mode the meter responds in 125ms Select the SLOW Mode meter responds in 1s to monitor a sound source that has a reasonably consistent noise level or t...

Страница 8: ...the arrow keys to set the correct month 4 Repeat step 3 to set the day hour minute and second 5 To confirm the new date and time press the DATE TIME button Note that the 407780 has a back up battery N...

Страница 9: ...th a dry cloth Do not use abrasives or solvents Analog output The 407780 has two analog output jacks located on its right side one for AC and one for DC For DC the meter transmits 10mv dB For AC the f...

Страница 10: ...and the measurement time in the supplied software The meter can now record remotely disconnected from the PC at the rate selected 4 Press the RECORD ERASE key to put the meter in Record Standby The RE...

Страница 11: ...ay 4dB per step over 100dB range with 50 segments Bargraph update rate 16 times per second Display warnings OVER and UNDER range status indicators Analog outputs DC 10mV per dB AC 2Vrms full scale Pow...

Страница 12: ...al or consequential damages FLIR s total liability is limited to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is express...

Отзывы: