Operation 

Connecting the Sensor 

1.  Remove the protective sheath from the sensor tip before use. Retain the sheath for 

sensor storage. 

2.  The sensor connects to the input jack at the top of the meter. The sensor jack is keyed, 

thereby allowing the sensor plug to be inserted from only one direction. Line up the 
plug with the jack before firmly inserting the sensor. Avoid undue force; forcing the 
sensor into the jack may bend the pins in the sensor plug. 

Powering the meter 

1.  Press the POWER button to turn the meter ON or OFF. If the meter does not switch on 

when the POWER button is pressed, check the battery. Refer to the battery installation 
and replacement section later in this manual. 

2.  This meter is equipped with an Automatic Power OFF feature. To conserve battery 

energy, the meter shuts off after 10 minutes. To defeat this feature, press the 
RECORD button to access the Record/Recall mode. In the Record/Recall mode, the 
meter will remain on until the user turns it off or until the battery weakens. 

LCD Contrast Adjustment 

It may be necessary to adjust the display contrast due to a change in viewing angle or 
voltage drift. Use the LCD Contrast adjustment wheel located on the right side of the meter 
to set the preferred contrast. 

Taking Measurements and using the RANGE select slide switch 

1.  Insert the sensor tip into the solution under test. 

2.  The conductivity and temperature readings will appear on the LCD. Conductivity 

measurements appear at the center of the LCD. Temperature readings appear on the 
bottom of the LCD in smaller digits. 

3.  Allow time for the measurements to stabilize. The LCD display digits will settle after a 

short stabilization period. 

4.  If dashes appear on the top of the LCD measurement area, the reading is over range. 

Slide the RANGE select switch to a higher range. If dashes appear on the bottom of 
the measurement display area, the reading is under range. Slide the RANGE switch to 
a lower range. 

Selecting 

o

C/

o

F units of measure 

Press the 

o

C/

o

F button to select the Temperature unit of measure. The display will reflect 

the currently selected unit of measure. 

Temperature Compensation using the FACTOR ADJ and TEMP buttons  

This meter has a user adjustable temperature compensation feature that accommodates a 
change in conductivity due to a change in solution temperature. The default compensation 
is 2% and applies to most common standard salt solutions. Other solutions have known 
temperature coefficients and the temperature adjustment allows the user to set it to match 
the solution. To change the coefficient: 

1.  Set the meter to the 200

µ

S range via the RANGE slide switch 

2.  Press the TEMP button 

3.  Press the Factor Adjust button to select the desired compensation factor (in increments 

of 0.1% per degree C). The display will show the factor as it is adjusted. 

4.  Once the desired factor is set, press the TEMP button. 

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

FAX 781.665.0780 - TestEquipmentDepot.com 

Содержание 407303

Страница 1: ...onductivity and Temperature a Conductivity Cell calibration adjustment temperature compensation and a RS 232 PC Interface This meter is shipped fully tested and calibrated and with proper use will provide years of reliable service 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com Back to the Extech 407303 Product Page ...

Страница 2: ...ings for later recall Auto Power OFF Automatic power off after 10 minutes PC Interface RS 232 serial data output Operating conditions Temperature 32 F to 122 F 0 C to 50 C 80 R H Power Supply 9V battery Power Consumption Approx 10 mA DC Weight 17 oz 482 g including batteries probe Dimensions Main instrument 7 1 x 2 8 x 1 3 180 x 72 x 32 mm Probe 0 87 22 mm diameter Measurement Range Resolution Acc...

Страница 3: ... the displayed reading C F Press to select the temperature unit of measure RECORD Press to begin tracking the MIN MAX readings RECALL Press to retrieve the MIN MAX readings FACTOR ADJUST Used with the TEMP button for temperature compensation RANGE Slide switch to manually select the measurement range TEMP Temperature coefficient compensation button 5 Battery compartment rear Remove protective hols...

Страница 4: ... temperature readings will appear on the LCD Conductivity measurements appear at the center of the LCD Temperature readings appear on the bottom of the LCD in smaller digits 3 Allow time for the measurements to stabilize The LCD display digits will settle after a short stabilization period 4 If dashes appear on the top of the LCD measurement area the reading is over range Slide the RANGE select sw...

Страница 5: ...n The REC MAX MIN and AVG icons should be off when the meter has exited the MIN MAX AVG Record Recall mode Conductivity Calibration 1 Keep the battery connected to the meter but remove it from the battery compartment 2 Prepare a standard conductivity solution of one or all of the following 200uS 2 000mS and 20 00mS 3 Place the probe into the calibration solution and adjust the corresponding potent...

Страница 6: ...ed at the top of the meter next to the sensor input jack The interface is intended for use with the data acquisition software and hardware kit For more information contact Extech or refer to the Software User Manual Battery Replacement The low battery indicator LBT appears on the LCD when the battery runs low To replace the battery 1 Remove the meter s rubber protective cover to access the rear ba...

Отзывы: