background image

Model 407114 Version 2.0 November 2005 

Air Flow Measurements (CMM / CFM) 

1.  Connect the sensor to the sensor input jack on top of the meter. 

2.  Turn on the meter using the Power button. 

3.  Select the FLOW mode using the FLOW/VELOCITY button. The LCD will display 

FLOW

 when the Flow function has been properly selected. 

4.  Select the desired air flow units: CMM (cubic meters per minute) or CFM (cubic feet 

per minute) using the UNIT button. The LCD will reflect the selection. 

5.  Press the AREA button to begin entering the area in m

2

 or ft

2

. Use the      button to 

increment the flashing digit, use the     button to decrement the flashing digit, use the       
button to select the next digit, and use the RECORD/RECALL    button to set the 
decimal point. Press the ENTER/RESET button when the area has been entered. The 
bottom display (normally used for the temperature display) will indicate the area 
entered in ft

2

 or m

2

. The main LCD displays the air flow in CFM (cubic feet per minute) 

or CMM (cubic meters per minute). If the CFM or CMM reading exceeds 9999, use the 
displayed X10 or X100 multiplier to calculate the reading. 

6.  For Air Flow measurements, three modes apply: The normal (default mode) where the 

actual flow is indicated, and the two modes described below: Press the FLOW/MODE 
button to select one of the two modes listed below: 

2/3V MAX MODE

: LCD displays 2/3 the measured value. 

AVG MODE

: Up to 20 readings can be taken separately and averaged. Select the AVG 

mode via the FLOW/MODE button and then press the AVG/START button to take a 
reading. Up to 20 readings can be taken and averaged. The lower LCD display 
provides a 1 to 20 counter and each time a reading is taken the counter is 
incremented. The main LCD displays the averaged air flow. 

7.  Note that the temperature function is not active in the Air Flow mode. 

Automatic Power OFF 

The Model 407114 turns off automatically after 15 minutes of operation. This feature was 
designed to conserve battery power. Note that when the meter is in the RECORD/RECALL 
mode, the Automatic Power OFF feature is disabled. 

Battery Replacement 

When 

LBT

 appears on the LCD, the 9V battery must be replaced. 

1.  Disconnect the sensor and the RS-232 PC Interface cable from the meter. 

2.  Remove the meter’s rubber protective jacket 

3.  Use a flat blade screwdriver or a coin to open the rear battery compartment 

4.  Replace the 9V battery 

5.  Close the battery compartment and replace the meter’s protective jacket 

RS-232 PC Interface 

The 407114 is equipped with a 3.5mm phono jack (meter top) for connection to a PC. To 
obtain the Model 407001 Windows

TM

 data acquisition kit, contact Extech Instruments. The 

kit includes a cable that connects the meter to a PC. Instructions for use are provided with 
the 407001 kit. 

 

Содержание 407114

Страница 1: ...User s Manual CFM Thermo Anemometer Air Velocity Air Flow and Temperature Model 407114 ...

Страница 2: ...atulations on your purchase of the Extech CFM Thermo Anemometer This instrument measures Air Velocity Air Flow volume and Temperature The RS 232 PC interface provides data acquisition functionality Proper use of this meter will provide years of reliable service ...

Страница 3: ...it data stream Operating Temperature 32 F to 122 F 0 C to 50 C Operating Humidity Max 80 RH Power Supply 9V battery Power Current Approx 8 3 mA DC Weight 0 84 lbs 381g including battery probe Dimensions Main instrument 7 1 x 2 8 x1 3 180 x 72 x 32mm Sensor Head 72mm Diameter Range Specifications Air Velocity Range Resolution Accuracy rdg m s meters per sec 0 40 25 00 m s 0 01 m s 2 0 2 m s km h ki...

Страница 4: ... this manual for more details This button also functions as the DECIMAL point when programming UNIT Press to select the unit of measure for Air Velocity This button also functions as the DOWN ARROW button when programming FLOW VELOCITY Press to select the mode of operation In FLOW mode the meter displays air volume In VELOCITY mode the meter displays air speed RIGHT Arrow button Used when programm...

Страница 5: ... display shows the temperature reading Data Hold Feature 1 While taking measurements press the HOLD button to freeze the LCD reading for later viewing 2 The DH indicator will appear on the LCD when the display is in Data Hold mode 3 Press HOLD again to return to normal operation Maximum and Minimum Recording The 407114 allows the user to record and view the highest MAX and lowest MIN readings 1 Pr...

Страница 6: ...button to select one of the two modes listed below 2 3V MAX MODE LCD displays 2 3 the measured value AVG MODE Up to 20 readings can be taken separately and averaged Select the AVG mode via the FLOW MODE button and then press the AVG START button to take a reading Up to 20 readings can be taken and averaged The lower LCD display provides a 1 to 20 counter and each time a reading is taken the counte...

Страница 7: ...e sender is responsible for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenance or repair or unauthorized modification Extech specifically disclaims any implied warranties or merchantability or fitness for ...

Страница 8: ...using the above formulae Unit of Measure Conversion Table m s ft min knots km h MPH 1 m s 1 196 87 1 944 3 6 2 24 1 ft min 0 00508 1 0 00987 0 01829 0 01138 1 knot 0 5144 101 27 1 1 8519 1 1523 1 km h 0 2778 54 69 0 54 1 0 6222 1 MPH 0 4464 87 89 0 8679 1 6071 1 Width W Height H Area A Width W x Height H CFM ft3 min Air Velocity ft min x Area ft2 CMM m3 min Air Velocity m sec x Area m2 x 60 Radius...

Отзывы: