Model 407114 Version 2.0 November 2005 

 Meter Description  

 

1.  Sensor plug. Can be inserted in the 

sensor jack from only one direction. 
Ensure proper orientation before 
exerting force to plug in the sensor. 

2.  PC Interface jack (RS-232). For 

use with the interface cable 
provided in the optional 407001 PC 
Data Acquisition kit. 

3.  LCD Display for Air Velocity, Air 

Flow, and Temperature. Units of 
measure and user alerts are also 
displayed on the LCD.  

4.  Vane Sensor. Hold the sensor in 

the flow of air to initiate a reading. 

5.  Protective rubber holster. Must be 

removed to access the rear battery 
compartment. 

6. Keypad 

(see 

below) 

 

POWER

 Press to turn the 

meter ON or OFF 

 

HOLD

 Press to freeze the 

displayed reading. Press again 
to unlock display 

 

C/F

 Press to select the unit of measure for Temperature readings 

 

RECORD/RECALL

     Used to store the highest, lowest, and average readings for 

later recall. Refer to the Operation section of this manual for more details. This 
button also functions as the DECIMAL point when programming. 

 

UNIT

     Press to select the unit of measure for Air Velocity. This button also 

functions as the DOWN ARROW button when programming 

 

FLOW/VELOCITY

 Press to select the mode of operation. In FLOW mode, the 

meter displays air 

volume

. In VELOCITY mode, the meter displays air 

speed

 

     RIGHT Arrow button. Used when programming. 

 

    UP Arrow button. Used when programming. 

 

FLOW MODE

 Press to select the Normal, AVG, or 2/3 MAX Air Flow operational 

mode. Refer to the Operation section of this manual for more details. 

 

AVG/START

 In the FLOW mode, press to store a reading in the AVERGGE utility. 

Up to 20 readings can be averaged. Refer to the Operation section of this manual 
for more details. 

 

ENTER/RESET

 Used in the Flow mode while entering the duct area. Refer to the 

Operation section of this manual. 

 

AREA

 Used when manually entering the area of a duct. Refer to the Operation 

section of this manual. 

7.  Battery compartment located on the back of the instrument. The rubber protective 

jacket must be removed from the meter to access the compartment. 

  
 
 
 

Содержание 407114

Страница 1: ...User s Manual CFM Thermo Anemometer Air Velocity Air Flow and Temperature Model 407114 ...

Страница 2: ...atulations on your purchase of the Extech CFM Thermo Anemometer This instrument measures Air Velocity Air Flow volume and Temperature The RS 232 PC interface provides data acquisition functionality Proper use of this meter will provide years of reliable service ...

Страница 3: ...it data stream Operating Temperature 32 F to 122 F 0 C to 50 C Operating Humidity Max 80 RH Power Supply 9V battery Power Current Approx 8 3 mA DC Weight 0 84 lbs 381g including battery probe Dimensions Main instrument 7 1 x 2 8 x1 3 180 x 72 x 32mm Sensor Head 72mm Diameter Range Specifications Air Velocity Range Resolution Accuracy rdg m s meters per sec 0 40 25 00 m s 0 01 m s 2 0 2 m s km h ki...

Страница 4: ... this manual for more details This button also functions as the DECIMAL point when programming UNIT Press to select the unit of measure for Air Velocity This button also functions as the DOWN ARROW button when programming FLOW VELOCITY Press to select the mode of operation In FLOW mode the meter displays air volume In VELOCITY mode the meter displays air speed RIGHT Arrow button Used when programm...

Страница 5: ... display shows the temperature reading Data Hold Feature 1 While taking measurements press the HOLD button to freeze the LCD reading for later viewing 2 The DH indicator will appear on the LCD when the display is in Data Hold mode 3 Press HOLD again to return to normal operation Maximum and Minimum Recording The 407114 allows the user to record and view the highest MAX and lowest MIN readings 1 Pr...

Страница 6: ...button to select one of the two modes listed below 2 3V MAX MODE LCD displays 2 3 the measured value AVG MODE Up to 20 readings can be taken separately and averaged Select the AVG mode via the FLOW MODE button and then press the AVG START button to take a reading Up to 20 readings can be taken and averaged The lower LCD display provides a 1 to 20 counter and each time a reading is taken the counte...

Страница 7: ...e sender is responsible for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenance or repair or unauthorized modification Extech specifically disclaims any implied warranties or merchantability or fitness for ...

Страница 8: ...using the above formulae Unit of Measure Conversion Table m s ft min knots km h MPH 1 m s 1 196 87 1 944 3 6 2 24 1 ft min 0 00508 1 0 00987 0 01829 0 01138 1 knot 0 5144 101 27 1 1 8519 1 1523 1 km h 0 2778 54 69 0 54 1 0 6222 1 MPH 0 4464 87 89 0 8679 1 6071 1 Width W Height H Area A Width W x Height H CFM ft3 min Air Velocity ft min x Area ft2 CMM m3 min Air Velocity m sec x Area m2 x 60 Radius...