![Extech Instruments 407113 Скачать руководство пользователя страница 7](http://html.mh-extra.com/html/extech-instruments/407113/407113_user-manual_2453110007.webp)
407113-EN v3.6 4/14
7
Battery Replacement
When the
LBT
icon appears on the LCD, the 9V battery must be replaced.
1. Remove the rubber holster that surrounds the entire meter
2. Slide off the rear battery compartment
3. Replace the 9V battery
4. Affix the battery compartment cover and the meter holster
Useful Equations and Conversions
Area equation for rectangular or square ducts
Area equation for circular ducts
Cubic equations
NOTE:
Measurements made in
inches
must be converted to
feet
or
meters
before using the above formulae.
Unit of Measure Conversion Table
m/s
ft/min
knots
km/h
MPH
1 m/s
1
196.87
1.944
3.6
2.24
1 ft/min
0.00508
1
0.00987
0.01829
0.01138
1 knot
0.5144
101.27
1
1.8519
1.1523
1 km/h
0.2778
54.69
0.54
1
0.6222
1 MPH
0.4464
87.89
0.8679
1.6071
1
Width (W)
Height (H)
Area (A) = Width (W) x Height (H)
CFM (ft
3
/min) = Air Velocity (ft/min) x Area (ft
2
)
CMM (m
3
/min) = Air Velocity (m/sec) x Area (m
2
) x 60
Radius
Area (A) = pi x r
2
Where pi = 3.14 and r
2
= radius x radius