Extech Instruments 407113 Скачать руководство пользователя страница 2

407113-EN v3.6   4/14 

2

Introduction 

Congratulations  on  your  purchase  of  the  Extech  407113  CFM  meter.  This  handheld  meter  can 
display Air Flow (CFM) with Area or Air Velocity with Temperature on a 10,000-count (0 to 9999) 
dual  display  LCD.  Other  functions  include  MIN/MAX  Record/Recall,  RS-232  PC  interface,  and 
Data Hold. The metal vane sensor offers rugged durability and higher temperature withstand than 
comparable devices. Careful use of this meter will provide years of reliable service. 

Specifications 

General Specifications 

Display  

 

Dual Display Multi-function 10,000-count (0 to 9999) LCD 

Measurements   

Air Velocity: m/s, km/h, ft/min, knots, mph; 
Air Flow: CMM (m

3

/min) and CFM (ft

3

/min); Temperature: °C and °F 

Data Hold 

 

Freezes displayed reading 

Sampling rate 

 

One (1) reading per second 

Sensors 

Air velocity/flow sensor: Metal angled vane arms with low-friction ball 
bearing. Temp. Sensor: Precision thermistor 

MIN/MAX 

Records/Recalls lowest and highest readings 

Auto Power OFF 

Automatic shut off after 15 minutes 

PC Interface 

RS-232 Serial Communications (16-bit data stream)  

Over range indication 

1___

 appears on the LCD 

Low battery indication 

LBT

 appears on the LCD 

Power supply 

 

9V Battery (consumption 8.3mA approx.) 

Operating conditions 

Meter: 32 to 122

o

F (0 to 50

o

C); 80% RH max. 

 

 

 

Sensor: 32 to 175

o

F (0 to 80

o

C) 

Dimensions / Weight 

Main instrument: 7.1 x 2.8 x1.3" (180 x 72 x 32mm) 
Sensor head: 2.8  (72mm) diameter 

Weight  

 

0.84 lbs. (381g) for meter and sensor 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание 407113

Страница 1: ...User Guide Heavy Duty CFM Thermo Anemometer Model 407113 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Страница 2: ...n and CFM ft 3 min Temperature C and F Data Hold Freezes displayed reading Sampling rate One 1 reading per second Sensors Air velocity flow sensor Metal angled vane arms with low friction ball bearing Temp Sensor Precision thermistor MIN MAX Records Recalls lowest and highest readings Auto Power OFF Automatic shut off after 15 minutes PC Interface RS 232 Serial Communications 16 bit data stream Ov...

Страница 3: ...iles per hour 1 1 78 3 mph 0 1 mph 2 0 4m h knots nautical miles per hour 1 0 to 68 knots 0 1 knots 2 0 4knots Air Flow Measurements Range Resolution Area CMM cubic meters per minute 0 999 900 m 3 min 0 001 to 100 0 to 9 999m 2 CFM cubic feet per minute 0 999 900 ft 3 min 0 001 to 100 0 to 9 999ft 2 Air Temperature Range Resolution Accuracy 32 to 175 o F 0 to 80 o C 0 1 o F C 1 5 o F 0 8 o C www G...

Страница 4: ...nterface jack 3 LCD Display 4 Vane 5 Rubber Holster 6 Keypad 7 Battery compartment rear Note To access the rear battery compartment first remove the rubber holster that jackets the meter www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Страница 5: ...when the display is in Data Hold mode 3 Press HOLD again to return to normal operation Maximum and Minimum Recording The 407113 allows the user to record and view the highest MAX and lowest MIN readings 1 Press the RECORD RECALL button once The REC indicator will appear on the display and the meter will begin recording the MAX and MIN values 2 Press the RECORD RECALL button to stop the MAX MIN rec...

Страница 6: ...e The temperature function is not active in the FLOW mode 2 3V MAX Flow Mode In this mode the meter will display two thirds the measured Flow Volume To access this mode ensure that the meter is in the FLOW mode via the FLOW VELOCITY button and then press the FLOW MODE button until the 2 3V MAX display icon appears Now the Flow display will be 2 3 the actual measurement To return to the normal disp...

Страница 7: ...ations NOTE Measurements made in inches must be converted to feet or meters before using the above formulae Unit of Measure Conversion Table m s ft min knots km h MPH 1 m s 1 196 87 1 944 3 6 2 24 1 ft min 0 00508 1 0 00987 0 01829 0 01138 1 knot 0 5144 101 27 1 1 8519 1 1523 1 km h 0 2778 54 69 0 54 1 0 6222 1 MPH 0 4464 87 89 0 8679 1 6071 1 Width W Height H Area A Width W x Height H CFM ft3 min...

Страница 8: ...r any direct indirect incidental or consequential damages FLIR s total liability is limited to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is expressed or implied Calibration Repair and Customer Care Services FLIR Systems Inc offers repair and calibration services for the Extech Instruments products we sell NIST certi...

Страница 9: ... conséquent direct indirect ou accidentel La responsabilité totale de FLIR est limitée à la réparation ou au remplacement du produit La garantie définie ci dessus est inclusive et aucune autre garantie écrite ou orale n est exprimée ou implicite Calibrage réparation et services après vente FLIR Systems Inc offre des services de calibrage et de réparation pour les produits Extech Instruments que no...

Страница 10: ...sito determinado y no será responsable por cualesquier daños directos indirectos incidentales o consecuentes La responsabilidad total de FLIR está limitada a la reparación o reemplazo del producto La garantía precedente es inclusiva y no hay otra garantía ya sea escrita u oral expresa o implícita Servicios de calibración reparación y atención a clientes FLIR Systems Inc ofrece servicios de reparac...

Отзывы: