3                            401020 and 401021      Ver. 1.03    11/99

 

4. OPERATION 

4.1 Insure that a fresh battery is installed in the adapter by moving the Power/Range 

Select switch from the OFF position to one of the range select positions. If the Power 
ON light doesn't glow or if the Low Battery LED does glow, check the battery. 

4.2 Connect the adapter to the COMMON and VOLTAGE inputs of a multimeter via the 

adapter's banana plugs. The polarity of the connection should be reversed if a negative 
light level reading is indicated on the multimeter. 

4.3 Set the mulitmeter to the mV or higher DC voltage position. Adjust the range to suit 

your display needs. Do not use autorange if possible. 

4.4 Select the maximum range via the Range Select Switch and move lower in range as 

needed. 

4.5 Hold the Light Sensor so that it faces the light source to be measured. The 

multimeter's display will indicate light level readings. Refer to Sections 5 and 6 for 
additional measurement considerations. 

5. CORRECTION FACTOR FOR DIFFERENT TYPES OF LIGHTING 

The light adapter is calibrated using a 
precise "Standard tungsten light source 
of 2856oK". If the meter is to measure 
a different type of lighting, the 
correction factor table, below, should 
be applied to the readings. Multiply the 
multimeter readings by the factors in 
the right column for a particular lighting 
type shown in the left column. 

6. RANGE SELECTION 

The meter has three measurement ranges (0-200, 0-2000, and 0-5000 Fc) or (0-2000, 0-
20000, and 0-50000) Lux). The proper range selection will produce the most accurate 
reading. Always select the range that produces the maximum number of digits without 
exceeding the maximum count for that particular range. For example, a reading of 1456 Fc 
should be read on the 0 - 2000 range, not the 0-5000 range. 

7. BATTERY REPLACEMENT 

When the Low Battery status LED glows, replace the battery as soon as possible. Open 
the Battery Compartment by removing the rear screw.  Remove the battery, replace it 
with a fresh 9V battery and re-assemble the meter. 

8. CALIBRATION / REPAIR SERVICES 

Extech offers complete repair and calibration services for all of the products we sell. For 
periodic calibration, NIST certification or repair of any Extech product, call customer 
service for details on services available. Extech recommends that calibration be performed 
on an annual basis to insure calibration integrity. 

9. WARRANTY 

EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and 
workmanship for one year from date of shipment (a six month limited warranty applies on sensors and 
cables). If it should become necessary to return the instrument for service during or beyond the warranty 
period, contact the Customer Service Department at (781) 890-7440 for authorization. A Return 
Authorization (RA) number must be issued before any product is returned to Extech. The sender is 
responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transit. 
This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, 
operation outside of specification, improper maintenance or repair, or unauthorized modification. Extech 
specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not 
be liable for any direct, indirect, incidental or consequential damages. Extech's total liability is limited to 
repair or replacement of the product. 
The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or 
implied. 

 

Lighting Type

Multiplication Factor 

MERCURY LAMP

 

x1.14

 

FLUORESCENT 
LAMP

 

x1.08

 

DAYLIGHT

 

x1.00

 

SODIUM

 

x1.22

 

METAL HALIDE

 

x1.00

 