background image

381235       Ver. 1.5    3/01

 

2

Specifications 

General Specifications 

 

Display 

0.5” (13mm) 8-digit LCD 

Functions 

Frequency, Data Hold, Relative, Max/Min/Avg & Period 

Input connectors 

BNC female 

Meter housing 

Durable ABS plastic 

Operating conditions 

32 to 122

o

F (0 to 50

o

C); < 90% RH 

Power supply 

4 x 1.5V AA cells 

105mA DC approx. 2500MHz/500MHz ranges 

Power consumption 

45mA DC approx. for 10MHz & Period ranges 

AC adapter (option) 

9VDC, 300 to 500mA (center socket pin positive) 

Dimensions; weight 

6.8 x 3.1 x 1.4” (173 x 80 x 35mm); 0.75 lbs (340g) 

Electrical Specifications 

Ch. A: 2500MHz 

Frequency range 

Ch. B: 500MHz 

Period range 

Ch. C: 10MHz 

Resolution, sample rate and 

sensitivity 

Refer to Sections 2.3 and 2.4 below  

Accuracy 

± (4 ppm + 1 digit) @ 23

o

C ±5

o

Time base 

4.194 MHz quartz crystal 

Temp. coefficient 

0.1 ppm per 

o

C typical 

5V peak to peak max. for 2500/500MHz range 

Input overload 

250V peak to peak max. for 10MHz & Period range 

Resolution and sampling rate specifications 

Range 

Gate time selection  Resolution (Hz)  Sampling time (sec) 

FAST 

1000 

0.5 

SLOW 

100 

2.75 

SLOW (select 1) 

200 

1.5 

2500MHz 

SLOW (select 2) 

500 

0.75 

FAST 

100 

0.75 

SLOW 

10 

SLOW (select 1) 

20 

500MHz 

SLOW (select 2) 

50 

1.5 

FAST 

10 

0.5 

SLOW 

1.25 

SLOW (select 1) 

0.2 

10MHz 

SLOW (select 2) 

0.1 

11 

NOTE:

 In SLOW Gate Time mode (Gate switch set to SLOW), pressing the RESO key 

scrolls the meter through SLOW, SLOW (Select1), and SLOW (Select 2) modes. 

Sensitivity specifications  

 100mVrms (50MHz to 75MHz) 

Ch.A: 2500MHz range 

 50mVrms (76MHz to 2500MHz) 

 120mVrms (10M to 35MHz) 

 50mVrms (36M to 350MHz) 

Ch.B: 500MHz range 

 120mVrms (351M to 450MHz) 

Ch.C: 10MHz range, Period 

 

50mVrms  

Notes on sensitivity: 

The apparent sensitivity when feeding a signal into Ch.A may vary 

considerably, particularly near 1GHz. The meter’s input impedance is a complex impedance, 

whereas the output of most signal generators is calibrated for a 50

 resistive load. This will 

result in a terminal voltage different to that expected by the generator, and therefore a 

different sensitivity. 

Содержание 381235

Страница 1: ...ntroduction Congratulations on your purchase of Extech s Frequency Counter This meter will measure frequency to 2 5GHz using test leads or the optional RF antenna This professional meter with proper care will provide years of safe reliable service ...

Страница 2: ...MHz Period range Resolution and sampling rate specifications Range Gate time selection Resolution Hz Sampling time sec FAST 1000 0 5 SLOW 100 2 75 SLOW select 1 200 1 5 2500MHz SLOW select 2 500 0 75 FAST 100 0 75 SLOW 10 6 SLOW select 1 20 5 500MHz SLOW select 2 50 1 5 FAST 10 0 5 SLOW 1 1 25 SLOW select 1 0 2 6 10MHz SLOW select 2 0 1 11 NOTE In SLOW Gate Time mode Gate switch set to SLOW pressi...

Страница 3: ...ress the Power OFF key 2 Slide the Range selector switch to desired measurement range Note that for the best accuracy and resolution be sure to use the most appropriate range 3 Apply the measurement signal to the input Channel A socket if the frequency of the signal is between 500MHz and 2500MHz 4 Apply the measurement signal to the input Channel B socket if the frequency of the signal is between ...

Страница 4: ...n alternately The 2nd CALL key press shows the minimum reading and the Lo icon alternately The 3rd CALL key press shows the average reading and the A icon alternately The Average reading will continue to update every 10 samples To return to normal operation press the RECORD key again Period Measurements Input the signal to Ch C Slide the Range selector switch to the PERIOD position Period measurem...

Страница 5: ...nsors and cables If it should become necessary to return the instrument for service during or beyond the warranty period contact the Customer Service Department at 781 890 7440 for authorization A Return Authorization RA number must be issued before any product is returned to Extech The sender is responsible for shipping charges freight insurance and proper packaging to prevent damage in transit T...

Отзывы: