Resistance Measurements 

WARNING: To avoid electric shock, disconnect power to the unit 
under test and discharge all capacitors before taking resistance 
measurements. 

1.

 

Before testing resistance, test the condition of the battery: 
a.

 

Set the function switch to ΩX1K or ΩX10 position. 

b.

 

Short the test leads. The needle deflects to the right side. 

c.

 

Keep the test leads shorted. Turn the 0 Ω adjustment 
dial (6) until the needle reads zero Ω.  

d.

 

If the needle will not zero, replace the battery. 

2.

 

Set the function switch to the X1K or X10 range. 

3.

 

Insert the black test lead into the COM input and the red 
test lead into the VmAΩ input. 

4.

 

Touch the test leads across the circuit. Disconnect one side 
of the part under test so the rest of the circuit will not 
interfere with the resistance reading.  

5.

 

Read the measured resistance on the green Ω scale and 
multiply the reading by 10 or 1000 (see step 2). 

6.

 

When switching between functions, always zero the needle 
again (step 1) to avoid inaccurate readings. 

Battery (1.5V and 9V) Testing 

1.

 

Set the function switch to the 1.5V or 9V position. 

2.

 

Insert the black test lead into the COM input and the red 
test lead into the VmAΩ input. 

3.

 

Touch the black test lead to the negative side of the battery 
and the red test lead to the positive side of the battery. 

4.

 

Use the display scale labeled BAT to determine the quality 
of the battery.  

Green=Good 

?=Replace Soon 

Red=Replace Immediately 

 
 
 

Shop for Extech products at:

1.888.610.7664

ShopExtech

.com

Содержание 38073A

Страница 1: ...USER MANUAL Mini Analog MultiMeter Model 38073A Please visit www extech com for user manual translations Shop for Extech products at 1 888 610 7664 ShopExtech com ...

Страница 2: ... equipment for industrial use with permanent connection to the fixed installation Safety Information To ensure the safe operation and service of the meter follow these instructions closely Failure to observe warnings can result in severe injury WARNINGS WARNINGS identify hazardous conditions and actions that could cause BODILY HARM or DEATH Keep hands and fingers behind the finger guards at all ti...

Страница 3: ...TIONS identify conditions and actions that could cause DAMAGE to the meter or equipment under test Disconnect test leads before changing the position of the rotary switch Do not expose the meter to extremes in temperature or humidity Safety Symbols This symbol adjacent to another symbol indicates the user must refer to the manual or user guide for further information Risk of electrical shock Equip...

Страница 4: ...tions WARNING Risk of electrocution High voltage circuits are extremely dangerous and should be measured with care Getting Started Before taking any measurements use a flat head screwdriver to turn the mechanical zero adjustment screw 2 until the pointer needle 7 points to the zeros at the left end of the scale For greater accuracy lay the meter on a flat non metallic surface when testing Shop for...

Страница 5: ... to read AC DC voltage measurements ACV Range DCV Range Scale Multiply reading by N A 2 5 0 250 0 01 10 10 0 10 1 50 50 0 50 1 250 250 0 250 1 300 300 0 300 1 DC mA Current Measurements 1 Set the function switch to the correct mA DC range 2 Remove power to the circuit under test and open the circuit at the point to be measured 3 Insert black test lead into the COM input and red test lead into the ...

Страница 6: ...nput 4 Touch the test leads across the circuit Disconnect one side of the part under test so the rest of the circuit will not interfere with the resistance reading 5 Read the measured resistance on the green Ω scale and multiply the reading by 10 or 1000 see step 2 6 When switching between functions always zero the needle again step 1 to avoid inaccurate readings Battery 1 5V and 9V Testing 1 Set ...

Страница 7: ...n the back of the meter 3 Open battery compartment replace 1 1 5V AA battery observing correct polarity Re assemble the meter before use Safety Never dispose of batteries in a fire If the meter is not to be used for 60 days or more remove the battery and store separately Never dispose of used batteries or rechargeable batteries in household waste As consumers users are legally required to take use...

Страница 8: ...y 10kΩ per 1VDC 4 5kΩ per 1VAC General Specifications Display Analog with Zero Adjustment Operating Temp Humidity 18 25 C 64 77 F 75 RH Operating Altitude 2000m 7000 ft maximum Battery One 1 1 5V AA Battery Fuse 500 mA 500V fast blow Dimensions 116 x 68 x 34 mm 4 6 x 2 7 x 1 3 Weight 120g 4 2 oz Safety Compliance EN61010 1 CATIII 300V Pollution Degree 2 Ingress Protection Rating IP20 Standards CE ...

Отзывы: