background image

Model 380652 Version 3.0 June 2005 

Operation 

Voltage Measurements 

1. Slide 

the 

Off/On/Hold

 switch to the 

ON

 position. 

2. Slide 

the 

V/A/

 switch to the 

V

 position. 

3.  Connect the red test lead to the 

V

 input terminal and the black test lead to the 

COM

 

input terminal. 

4.  If the voltage to be measured is AC, slide the 

AC/DC

 switch to the 

AC

 position. 

5.  If the voltage to be measured is DC , slide the 

AC/DC

 switch to the 

DC

 position 

6.  Connect test lead probes to the circuit under test. 
7.  Read voltage values on the LCD. 

Resistance Measurements 

1. Slide 

the 

Off/On/Hold

 switch to the 

ON

 position. 

2. Slide 

the 

V/A/

 switch to the 

 position. 

3.  Connect the red test lead to the 

Ω 

input terminal and the black test lead to the 

COM

 

input terminal. 

4.  If the resistance being measured is in a circuit, remove all power to that circuit and 

discharge all capacitors. 

5.  Connect test lead probes to the circuit (resistance) under test. 
6.  Read the resistance values on the LCD. 

Note: There are small stray resistances (approx. 0.8 ohms) in the test leads. Before 
taking measurements short the test leads and note the meter reading. After taking a 
measurement, deduct the stray resistance value (noted earlier) from the displayed 
reading. 

AC Current Measurements 

1. Slide 

the 

Off/On/Hold

 switch to the 

ON

 position. 

2. Slide 

the 

V/A/

 switch to the 

A

 position. 

3. Slide 

the 

AC/DC

 switch to the 

AC

 position. 

4.  Determine the highest anticipated current (200A, 20A) 

and set the 

20A/200A

 switch to the appropriate position. 

5.  Press the trigger to open the current jaw. Clamp around 

the conductor under test. 

6.  Read ACA values on the digital display. 

DC Current Measurements 

1. Slide 

the 

Off/On/Hold

 switch to the 

ON

 position. 

2.  Slide the V/A/

 switch to the 

A

 position 

3. Slide 

the 

AC/DC

 switch to the 

DC

 position. 

4. Slide 

the 

20A/200A

 switch to the 

200A

 position. 

5.  Adjust the DCA Zero knob for a zero reading.  
6.  Press the trigger to open the current jaw. Clamp around the conductor under test. 
7.  Read DCA values on the digital display. 

Содержание 380652

Страница 1: ...and displays Voltage Current and Resistance The DCA zero adjustment permits relative and offset DC current measurements Careful use of this meter will provide years of reliable service 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com Back to the Extech 380652 Product Page ...

Страница 2: ...air or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is expressed or implied Specifications General Specifications Display 3 digit 1999 display counts LCD Measurements AC DC Voltage AC DC Current and Resistance Polarity Automatic Switching indicates negative polarity Current Sensor Hall effect sensor Zero Adjustment Adjustment di...

Страница 3: ... connected correctly to its snap terminal and placed in the battery compartment 2 Slide the Off On Hold switch to the ON position except when using the DATA HOLD function 3 Place the RED test lead into the proper input terminal before making measurements 4 Remove the test leads from the circuit under test when changing the measurement range 5 Do not exceed the maximum rated input voltage 6 Always ...

Страница 4: ... resistance under test 6 Read the resistance values on the LCD Note There are small stray resistances approx 0 8 ohms in the test leads Before taking measurements short the test leads and note the meter reading After taking a measurement deduct the stray resistance value noted earlier from the displayed reading AC Current Measurements 1 Slide the Off On Hold switch to the ON position 2 Slide the V...

Страница 5: ...able operating voltage approx 7V the LO BAT icon appears on the LCD display This indicates that the battery needs to be replaced soon However in spec measurements can still be made for several hours after the low battery indicator appears To replace the battery 1 Open the battery compartment with a screwdriver or small coin 2 Replace the 9V battery and secure the compartment cover Cleaning Use onl...

Отзывы: