Extech Instruments 380580 Скачать руководство пользователя страница 3

380580-en-US_v2.6   6/20 

Preliminary Checks 

Current Regulation Check 

1.  Connect the current leads C

1

 and C

2

 to the meter. 

2.  Set the function switch to the 200.0mΩ range. 
3.  Momentarily press the 

TEST R

P

 button. The meter will 

intermittently beep and R

will light 

4.  Short the current leads C

1

 to C

2.

 

5.  The R

C

 LED should go off, indicating that the meter is 

operating correctly. 

6.  Momentarily press the 

TEST R

P

 button to stop the test 

7.  The meter will return to 

NO TEST 

status. 

Voltage Measurement check 

1.  With the current test leads C

1

 and C

shorted, connect and 

short the potential (voltage) leads P

1

 and P

2

2.  Set the Function switch to the 200.0mΩ position. The 

NO 

TEST 

status LED will light. 

3.  Momentarily press the 

TEST R

P

 button. (the meter will 

intermittently beep) 

4.  The display should indicate 00.0 
5.  Momentarily press  the 

TEST R

P

 button to stop the test. 

The 

NO TEST 

status LED will light. 

6.  Remove the shorts from P

1

 and P

2

, and C

1

 and C

2

 and

 

 

7.  Short the test leads P

1

 to C

1

 and P

2

 to C

2

 

8.  The R

P

 LED as well as the 

NO TEST 

status LED should 

light indicating an over-voltage or over-range 

9.  Turn the rotary selector switch to OFF 

Polarity check 

1.  Short the test leads P

1

 to C

2

 and P

2

 to C

1

 together. 

2.  Set the Function switch to the 200.0mΩ position. 
3.  The “

-1

“ negative indicator should appear in the display. 

Operation check 

Use the Kelvin clips for this test 

1.  Short the all the leads (P

1

, P

2

 ,C

2

,C

1

) together. 

The P

1

P

2

 ,C

2

,C

order is important.

 

2.  Set the Function switch to the 200.0mΩ position. The 

NO TEST 

status LED will light. 

3.  Momentarily press the 

TEST R

P

 button. (the meter will 

intermittently beep  

4.  The display should indicate near 00.0 (+/-0.2mΩ) 

depending on the test clip connections and both R

P

 

and R

C

 LEDs should remain off. 

5.  Momentarily press the 

TEST R

P

 button to stop the test. 

Note:

 These tests can be performed on any range. 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание 380580

Страница 1: ...USER GUIDE Model 380580 Battery Powered Milliohm Meter www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Страница 2: ... 2 to Kelvin Clip Red P1 Green C1 Black P2 and Blue C2 Measurement Considerations 1 Do not apply voltage to the meter input terminals Meter damage may result 2 Always insure that the circuit to be measured is switched OFF isolated and completely de energized before connecting the test leads 3 If the Over Temperature LED NO TEST indicator is lit allow the instrument to cool down before proceeding f...

Страница 3: ...e NO TEST status LED will light 6 Remove the shorts from P1 and P2 and C1 and C2 and 7 Short the test leads P1 to C1 and P2 to C2 8 The RP LED as well as the NO TEST status LED should light indicating an over voltage or over range 9 Turn the rotary selector switch to OFF Polarity check 1 Short the test leads P1 to C2 and P2 to C1 together 2 Set the Function switch to the 200 0mΩ position 3 The 1 n...

Страница 4: ...f the potential test leads as shown in the diagram below 3 For a short test duration of 10 seconds press the TEST RP button for less than 2 seconds This EnerSave feature can be used to conserve battery power 4 For a long test duration of 60 seconds press the TEST RP button for more than 3 seconds 5 During the test the meter will intermittently beep At the end of the test or if the test is stopped ...

Страница 5: ...e resistance of the device under test Thermal Effects Temperature can have a significant effect on the performance of milliohmeter due to the temperature coefficient of the resistance under test and thermal EMF s across dissimilar conductors Most conductors have a large temperature coefficient of resistance For example 0 4 C for copper A copper conductor that has a resistance of 10 00m ohm at 20 C...

Страница 6: ...the two screws to open the battery compartment 3 Always use a fuse of the proper size and value Current Circuit Fuse 500mA 250V 5x20mm fast blow F2 1 Fuse protection for the current terminals 2 If the fuse is blown the RC LED will stay on 3 The fuse is located under the printed circuit board 4 There are 4 mounting screws that have to be removed 5 Two screws are located under the black feet on the ...

Страница 7: ...10 mm with cover Fuses 3 fuses all are 500mA 250V 5x20mm fast blow Range Specifications Range Resolution Test Current Accuracy reading Open Circuit Voltage 200 0m 0 1m 100mA 0 5 2 digits 4 2V 2000m 1m 100mA 0 5 2 digits 4 2V 20 00 0 01 10mA 0 5 2 digits 4 2V 200 0 0 1 10mA 0 5 2 digits 4 3V 2000 1 1mA 0 5 2 digits 4 4V International Symbols This symbol adjacent to another symbol or terminal indica...

Отзывы: