Introduction 

Congratulations on your purchase of Extech’s Model 380580 Battery Powered Milliohm Meter. This 

device offers five resistance ranges with resolution as low as 0.1m

. The 4-wire Kelvin clip connection 

ensures optimum accuracy. Typical applications include transformer, motor coil, and PC Board 

resistance measurements. This professional meter, with proper care, will provide years of safe reliable 

service. 

Meter Description 

1.  Current Terminals 
2.  Potential Measurement Terminals 
3.  LCD Display 
4.  Start/Stop Test Button 
5.  Range Select/Power Switch 
6.  LED Error Lights 

• No Test/Over Temperature 
• R

P

 Voltage Regulation 

• R

C

 Current Regulation 

7.  Current/Range Table 

Leads 

Current Leads- Banana plug to alligator clip 
 

C1- Green 

 

C2- Blue 

Voltage Potential Leads- Banana plug to alligator clip 
 

P1- Red 

 

P2- Black 

Kelvin Clips- Banana Plugs (2) to Kelvin Clip 
 

Red (P1) Green (C1) 
Black (P2) and Blue (C2) 

Measurement Considerations 

1.  Do not apply voltage to the meter input terminals. Meter damage may result. 
2.  Always insure that the circuit to be measured is switched OFF, isolated and completely de-

energized before connecting the test leads. 

3.  If the Over Temperature LED (NO TEST) indicator is lit, allow the instrument to cool down 

before proceeding further. 

4.  The R

C

 led indicates when the test current falls out of regulation. Selecting a higher range 

may eliminate the condition. 

5.  The R

P

 led indicates when the voltage on the device under test is too high. Selecting a lower 

range may eliminate the condition. 

6.  If either the R

C

 or R

P

 led is on, the measurement may be in error. 

7.  The current terminals are fuse protected. 
8.  Keep the potential test leads as short as possible. Long leads may introduce noise. 
9.  When using the four separate alligator clip leads always place the current leads outside the 

potential leads. 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание 380580

Страница 1: ...USER GUIDE Model 380580 Battery Powered Milliohm Meter www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Страница 2: ... 2 to Kelvin Clip Red P1 Green C1 Black P2 and Blue C2 Measurement Considerations 1 Do not apply voltage to the meter input terminals Meter damage may result 2 Always insure that the circuit to be measured is switched OFF isolated and completely de energized before connecting the test leads 3 If the Over Temperature LED NO TEST indicator is lit allow the instrument to cool down before proceeding f...

Страница 3: ...e NO TEST status LED will light 6 Remove the shorts from P1 and P2 and C1 and C2 and 7 Short the test leads P1 to C1 and P2 to C2 8 The RP LED as well as the NO TEST status LED should light indicating an over voltage or over range 9 Turn the rotary selector switch to OFF Polarity check 1 Short the test leads P1 to C2 and P2 to C1 together 2 Set the Function switch to the 200 0mΩ position 3 The 1 n...

Страница 4: ...f the potential test leads as shown in the diagram below 3 For a short test duration of 10 seconds press the TEST RP button for less than 2 seconds This EnerSave feature can be used to conserve battery power 4 For a long test duration of 60 seconds press the TEST RP button for more than 3 seconds 5 During the test the meter will intermittently beep At the end of the test or if the test is stopped ...

Страница 5: ...e resistance of the device under test Thermal Effects Temperature can have a significant effect on the performance of milliohmeter due to the temperature coefficient of the resistance under test and thermal EMF s across dissimilar conductors Most conductors have a large temperature coefficient of resistance For example 0 4 C for copper A copper conductor that has a resistance of 10 00m ohm at 20 C...

Страница 6: ...the two screws to open the battery compartment 3 Always use a fuse of the proper size and value Current Circuit Fuse 500mA 250V 5x20mm fast blow F2 1 Fuse protection for the current terminals 2 If the fuse is blown the RC LED will stay on 3 The fuse is located under the printed circuit board 4 There are 4 mounting screws that have to be removed 5 Two screws are located under the black feet on the ...

Страница 7: ...10 mm with cover Fuses 3 fuses all are 500mA 250V 5x20mm fast blow Range Specifications Range Resolution Test Current Accuracy reading Open Circuit Voltage 200 0m 0 1m 100mA 0 5 2 digits 4 2V 2000m 1m 100mA 0 5 2 digits 4 2V 20 00 0 01 10mA 0 5 2 digits 4 2V 200 0 0 1 10mA 0 5 2 digits 4 3V 2000 1 1mA 0 5 2 digits 4 4V International Symbols This symbol adjacent to another symbol or terminal indica...

Отзывы: