Specifications 

General Specifications 

Circuit

 

Custom one-chip LSI microprocessor circuit

 

Display 

 

20 mm (0.8") 2000 Count LED Display

 

Connection type

 

4-Terminal Kelvin

 

Measurement ranges 

Seven ranges (see listing below)

 

Test voltage 

5V DC 

Comparator 

Built-in HI/LO/GO testing with audible beeper 

Measurement speed 

Resistance 1 second, comparator mode 0.5 seconds 

Zero adjust 

Automatic (no adjustment necessary) 

Over-range indication

 

Display reads "1____" when the reading is out of range

 

Operating Temperature 

 

0°C to 50°C (32°F to 122°F)

 

Operating Humidity

 

Max. 80% RH

 

Power Supply

 

110V (380560) or 220V (380562) ±15%, 50/60Hz  

 

Weight      

 

2.2kg (4.85 lbs)

 

Dimensions   

 

280 x 210 x 90mm (11 x 8.3 x 3.5")

 

Replacement Kelvin leads 

380465 

Range Specifications 

Range

 

Resolution

 

Test Current 

Accuracy (%rdg)

 

Test Voltage

 

20 m

 

0.01m

 

1A

 

± (0.2% + 6 digits)

 

2.7V DC 

200 m

 

0.1m

 

1A

 

± (0.2% + 4 digits) 

3.3V DC 

 

.001

 

0.1A

 

3.5V DC 

20 

 

.01

 

10mA

 

4.1V DC 

200 

 

0.1 

 

1mA

 

4.5V DC 

2K 

 

.001k

 

0.1mA

 

20K 

 

.01k

 

10uA

 

Note: Specifications based on RF Field Strength <3V/m and frequency <30MHz 

 

International Symbols 

 

 

 

 

Содержание 380560

Страница 1: ...User Manual High Resolution Benchtop MilliOhm Meter Models 380560 and 380562 Additional User Manual Translations available at www extech com ...

Страница 2: ...istance measurements Careful use of this meter will provide years of reliable service Meter Description 1 LED Display 2 HI LO GO Status Indicators 3 Power and Range pushbuttons 4 Kelvin clip leads 5 SET TEST and HI LO pushbuttons 6 Comparator adjustment screws 7 Kelvin clip lead input terminals Note The power cable input and audible alert button are on the rear of the instrument Measurement Precau...

Страница 3: ...O HI pushbutton to HI 5 Adjust the HI comparator screw to the desired high limit 6 Position the SET TEST pushbutton to TEST 7 Set the rear buzzer switch ON or OFF When ON the meter will sound an audible tone for each GO measurement Each time a measurement is taken the appropriate Comparator status LED will light If the measurement is lower than the LO setting the LO LED will light If the measureme...

Страница 4: ...ge indication Display reads 1____ when the reading is out of range Operating Temperature 0 C to 50 C 32 F to 122 F Operating Humidity Max 80 RH Power Supply 110V 380560 or 220V 380562 15 50 60Hz Weight 2 2kg 4 85 lbs Dimensions 280 x 210 x 90mm 11 x 8 3 x 3 5 Replacement Kelvin leads 380465 Range Specifications Range Resolution Test Current Accuracy rdg Test Voltage 20 mΩ 0 01mΩ 1A 0 2 6 digits 2 ...

Страница 5: ...libration for most of our products Contact us for information on calibration and repair availability refer to the contact information below Annual calibrations should be performed to verify meter performance and accuracy Product specifications are subject to change without notice Please visit our website for the most up to date product information www extech com Contact Customer Support Customer S...

Отзывы: